Kora Y, Suzuki Y, Inatomi M, Ozawa T, Fukado Y
Department of Ophthalmology, Showa University, Tokyo, Japan.
Ophthalmic Surg. 1990 Apr;21(4):266-71.
We describe our modification of the most accurate intraocular lens power calculation formula currently used, the SRK formula, to improve the accuracy with which it can be used to calculate the power of IOLs that are implanted in severely myopic eyes, especially in those with exceptional axial length. This modified formula, which we call the L-SRK, is I = A - 2.5 L - 0.9 K - 1.69R - 1.69 (where I = the actual implanted IOL power; A = the A-constant; L = the axial length; K = the average keratometer reading; and R = the predicted postoperative refraction). The results achieved using this modified formula demonstrate its superior accuracy in calculating lens powers for severely myopic eyes.