School of Engineering, University of Waikato, Private Bag 3105, Hamilton 3240, New Zealand.
Phys Med Biol. 2013 Jun 7;58(11):3599-613. doi: 10.1088/0031-9155/58/11/3599. Epub 2013 May 3.
The electrical conductivity of small samples of mouse cortex (in vitro) has been measured at 10 kHz through the four-electrode method of van der Pauw. Brain slices from three mice were prepared under seizing and non-seizing conditions by changing the concentration of magnesium in the artificial cerebrospinal fluid used to maintain the tissue. These slices provided 121 square samples of cortical tissue; the conductivity of these samples was measured with an Agilent E4980A four-point impedance monitor. Of these, 73 samples were considered acceptable on the grounds of having good electrical contact between electrodes and tissue excluding outlier measurements. Results show that there is a significant difference (p = 0.03) in the conductivities of the samples under the two conditions. The seizing and non-seizing samples have mean conductivities of 0.33 and 0.36 S m(-1), respectively; however, these quantitative values should be used with caution as they are both subject to similar systematic uncertainties due to non-ideal temperature conditions and non-ideal placement of electrodes. We hypothesize that the difference between them, which is more robust to uncertainty, is due to the changing gap junction connectivity during seizures.
通过范德堡(van der Pauw)的四电极法,我们测量了 10 kHz 下小样本鼠皮层(体外)的电导率。在利用人工脑脊液维持组织的过程中,通过改变镁的浓度,从三只老鼠的大脑切片中制备了处于抽搐和非抽搐状态下的脑切片。这些切片提供了 121 个皮质组织的方块样本;使用安捷伦 E4980A 四点阻抗监测器测量了这些样本的电导率。其中,有 73 个样本被认为具有良好的电极与组织之间的电接触,且排除了异常值测量,因此是可以接受的。结果表明,在两种条件下,样本的电导率存在显著差异(p = 0.03)。抽搐和非抽搐样本的平均电导率分别为 0.33 和 0.36 S m(-1);然而,由于非理想的温度条件和电极位置不理想,这些定量值都存在类似的系统不确定性,因此应谨慎使用。我们假设,它们之间的差异更能抵抗不确定性,这是由于在抽搐期间缝隙连接连通性的变化所致。