J Cataract Refract Surg. 2013 Jun;39(6):888-97. doi: 10.1016/j.jcrs.2013.01.035.
To compare intraocular pressure (IOP) measurements using a new tonometer-pachymeter device (Tonopachy), Goldmann applanation tonometry (GAT), dynamic contour tonometry (DCT), and noncontact tonometry (NCT) before and after myopic laser-assisted subepithelial keratectomy (LASEK).
Institute of Vision Research, Department of Ophthalmology, Yonsei University College of Medicine, Seoul, South Korea.
Prospective comparative observational study.
Nine IOP values were measured using 4 tonometers (6 IOPs by tonometer-pachymeter and 1 each by GAT, DCT, and NCT) and compared preoperatively and 3 months postoperatively. Of the 6 IOP values measured by the tonometer-pachymeter, 1 showed uncorrected IOP and 5 indicated central corneal thickness (CCT)-corrected IOPs through different formulas.
Preoperatively and postoperatively, the means of 9 IOP values measured by 4 tonometers were statistically different. The IOP measurements by the tonometer-pachymeter correlated with all other IOP values preoperatively; however, there was no correlation with IOP measurements using DCT postoperatively. Some IOPs using the tonometer-pachymeter were interchangeable with those using GAT preoperatively, but not with DCT preoperatively or postoperatively. The corrected IOP values of the tonometer-pachymeter formula 3 and DCT did not change after surgery. The percentage change in CCT and corneal curvature and change in diopters correlated with the percentage change in IOP measurements by GAT and NCT but not with those using DCT and the tonometer-pachymeter.
Among the 9 IOP values, corrected IOP using the tonometer-pachymeter formula 3 showed similar IOP values after LASEK, as did DCT.
No author has a financial or proprietary interest in any material or method mentioned.
比较新眼压计-角膜测厚仪(Tonopachy)、压平眼压计(GAT)、动态轮廓眼压计(DCT)和非接触眼压计(NCT)在近视激光辅助上皮下角膜切除术(LASEK)前后的眼压(IOP)测量值。
韩国延世大学医学院视觉研究所眼科。
前瞻性比较观察研究。
使用 4 种眼压计(Tonopachy 测量 6 个 IOP,GAT、DCT 和 NCT 各测量 1 个)测量 9 个 IOP 值,并分别在术前和术后 3 个月进行比较。在 Tonopachy 测量的 6 个 IOP 值中,1 个为未矫正 IOP,5 个通过不同公式表示中央角膜厚度(CCT)校正的 IOP。
术前和术后,4 种眼压计测量的 9 个 IOP 值平均值存在统计学差异。Tonopachy 测量的 IOP 与术前所有其他 IOP 值均相关;然而,与术后 DCT 测量的 IOP 无相关性。Tonopachy 术前与 GAT 测量的部分 IOP 值可互换,但与术前或术后 DCT 测量的 IOP 值不可互换。Tonopachy 公式 3 的校正 IOP 值和 DCT 的校正 IOP 值在手术后均未发生变化。CCT 和角膜曲率的百分比变化以及屈光度的变化与 GAT 和 NCT 测量的 IOP 值百分比变化相关,但与 DCT 和 Tonopachy 测量的 IOP 值百分比变化无关。
在 9 个 IOP 值中,Tonopachy 公式 3 的校正 IOP 值与 DCT 相似,术后 LASEK 后 IOP 值也相似。