婴幼儿带套囊气管内导管:一种持续测量套囊内压力的技术。

Cuffed endotracheal tubes in infants and children: a technique to continuously measure the intracuff pressure.

作者信息

Gopalakrishnan Senthil, Barry N'Diris, Rice Julie, Tobias Joseph D

机构信息

Department of Anesthesiology & Pain Medicine, Nationwide Children's Hospital, Columbus, OH, United States.

出版信息

Int J Pediatr Otorhinolaryngol. 2013 Jul;77(7):1135-8. doi: 10.1016/j.ijporl.2013.04.020. Epub 2013 May 18.

Abstract

OBJECTIVES

Various devices have been used to measure the intracuff pressure (CP) of an endotracheal tube at the time of inflation; however, no device has found widespread acceptance for the continuous monitoring of CP. We devised a simple method to continuously measure the CP using an invasive pressure monitoring setup (IPMS), which is used routinely in the operating room to monitor arterial or central venous pressures. The accuracy of the device was compared to those obtained from a commercially available and clinically used manometer (MM).

METHODS

Size 4.0, 5.0 and 6.0 mm ID cETTs were placed into one of 3 sizes of polyvinylchloride (PVC) tubes. The cuff of the cETT was inflated after inserting the cETT into the PVC pipes. After inflation, the CP was simultaneously checked using the MM and the IPMS. A total of 100 simultaneous (IPMS and MM) readings were obtained from each of the 3 sizes of cETT with the intracuff pressure randomly varying between 10 and 40 cmH₂O. Statistical analysis included a Bland-Altman comparison to determine the bias, mean, and 95% levels of agreement (LOA), and a linear regression analysis.

RESULTS

Linear regression analysis demonstrated an R(2) value of 0.988, 0.9899, and 0.9879 when comparing the pressure from the IPMS and MM from the 4.0, 5.0 and 6.0 mm cETT, respectively.

CONCLUSIONS

A standard IPMS can be used to continuously monitor intracuff pressure with the use of a cETT. Although various techniques can be used safely to avoid hyperinflation of the cuff, they provide only an instantaneous measure of the CP. The current technique allows for continuous monitoring of CP during prolonged cases or in other clinical scenarios when such monitoring is required.

摘要

目的

在气管内导管充气时,已使用多种设备来测量其套囊内压力(CP);然而,尚无设备获得广泛认可用于持续监测CP。我们设计了一种使用有创压力监测装置(IPMS)持续测量CP的简单方法,该装置在手术室中常用于监测动脉或中心静脉压力。将该装置的准确性与从市售且临床使用的压力计(MM)获得的准确性进行比较。

方法

将内径为4.0、5.0和6.0 mm的气管内导管(cETT)放入3种尺寸的聚氯乙烯(PVC)管之一中。将cETT插入PVC管后,对其套囊进行充气。充气后,使用MM和IPMS同时检查CP。从3种尺寸的cETT中每种随机获取100次(IPMS和MM)同步读数,套囊内压力在10至40 cmH₂O之间随机变化。统计分析包括Bland - Altman比较以确定偏差、均值和95%一致性界限(LOA),以及线性回归分析。

结果

分别比较4.0、5.0和6.0 mm cETT的IPMS和MM压力时,线性回归分析显示R²值分别为0.988、0.9899和0.9879。

结论

标准的IPMS可用于通过cETT持续监测套囊内压力。尽管可安全使用各种技术以避免套囊过度充气,但它们仅提供CP的瞬时测量值。当前技术允许在长时间手术或其他需要此类监测的临床场景中持续监测CP。