Department of Anesthesiology and Critical Care Medicine, CHU de Caen, Caen, F-14000, France; Univ Caen, Faculty of Medicine EA4650, Caen, F-14000, France.
J Crit Care. 2013 Oct;28(5):881.e1-6. doi: 10.1016/j.jcrc.2013.05.004. Epub 2013 Jun 13.
The present study aimed to compare peripheral regional tissue oxygen saturation (rSO2) values and desaturation/resaturation rates given by INVOS and EQUANOX devices.
Twenty healthy volunteers were investigated during 4 experimental steps: baseline, hyperoxia, ischemia, and reperfusion. For each volunteer, 2 sensors INVOS and 2 sensors EQUANOX were placed on both left and right calves. Blood pressure, heart rate, and peripheral pulse oximetry were monitored.
Peripheral rSO2 ranged from 40% to 95% (INVOS) and from 47% to 100% (EQUANOX): 81±12 vs 82±9 (P=.469). A significant relationship was found at baseline between absolute values of INVOS and EQUANOX (n=40; R2=0.159; P=.011). Bias was -0.4%, and limits of agreement were ±15.1%. The percent maximum differences vs baseline values during dynamic maneuvers were 33%±19% (95% confidence interval, 24-42) and 21%±14% (95% confidence interval, 15-28) for INVOS and EQUANOX, respectively. No significant relationship was observed between percent maximum differences in INVOS and EQUANOX (n=20; R2=0.128; P=.122). Rates of desaturation/resaturation during occlusive vascular tests were 3.65% per minute vs 2.36% per minute (P=.027) and 30.42% per minute vs 16.28% per minute (P=.004) for INVOS and EQUANOX, respectively.
INVOS and EQUANOX are not comparable in measuring both absolute values and dynamic changes of peripheral rSO2 and near-infrared spectroscopy-derived parameters during occlusion vascular tests.
本研究旨在比较 INVOS 和 EQUANOX 设备测量的外周组织氧饱和度(rSO2)值和饱和度/复饱和度率。
在 4 个实验步骤中对 20 名健康志愿者进行了研究:基础状态、高氧、缺血和再灌注。每位志愿者的左右小腿上放置 2 个 INVOS 传感器和 2 个 EQUANOX 传感器。监测血压、心率和外周脉搏血氧饱和度。
外周 rSO2 范围为 40%至 95%(INVOS)和 47%至 100%(EQUANOX):81±12 与 82±9(P=.469)。在基础状态下,INVOS 和 EQUANOX 的绝对值之间存在显著关系(n=40;R2=0.159;P=.011)。偏倚为-0.4%,一致性界限为±15.1%。在动态操作期间,与基线值相比,最大差异百分比分别为 INVOS 33%±19%(95%置信区间,24-42)和 EQUANOX 21%±14%(95%置信区间,15-28)。INVOS 和 EQUANOX 的最大差异百分比之间未观察到显著关系(n=20;R2=0.128;P=.122)。在闭塞性血管试验中,饱和度/复饱和度率分别为 INVOS 3.65%/分钟和 EQUANOX 2.36%/分钟(P=.027)以及 INVOS 30.42%/分钟和 EQUANOX 16.28%/分钟(P=.004)。
在测量闭塞性血管试验中周围 rSO2 的绝对值和动态变化以及近红外光谱衍生参数时,INVOS 和 EQUANOX 不可比。