Department of Cardiothoracic surgery, St Thomas' hospital, London, United Kingdom.
Catheter Cardiovasc Interv. 2014 Jan 1;83(1):E150-4. doi: 10.1002/ccd.25076. Epub 2013 Aug 12.
We report two cases of delayed migration of the Sapien XT device after a successful mitral valve-in-valve (VIV) implantation. The procedure was performed through a transapical approach. Echocardiography was used to choose the size of the Sapien XT device. Although the immediate results were satisfactory both the cases presented with severe regurgitation (1 week and 3 months). Investigations revealed atrial migration of the Sapien device in both the cases, which was confirmed at the time of reoperation. We discuss possible mechanisms, which could have resulted in the delayed migration and highlight the difference between VIV procedures in mitral position versus other positions.
我们报告了两例成功的二尖瓣经皮介入置换术(VIV)后,Sapien XT 装置延迟移位的病例。该手术通过经心尖途径进行。术中采用超声心动图选择 Sapien XT 装置的大小。尽管即刻结果令人满意,但两例患者均出现严重反流(分别为术后 1 周和 3 个月)。检查发现两例患者的 Sapien 装置均发生心房移位,在再次手术时得到证实。我们讨论了可能导致延迟移位的机制,并强调了二尖瓣位置与其他位置的 VIV 手术之间的差异。