使用MOSFET剂量仪评估用于膝部区域的四肢CBCT、多层螺旋CT(MSCT)和传统X射线的有效辐射剂量。

Assessment of effective radiation dose of an extremity CBCT, MSCT and conventional X ray for knee area using MOSFET dosemeters.

作者信息

Koivisto Juha, Kiljunen Timo, Wolff Jan, Kortesniemi Mika

机构信息

Department of Physics, University of Helsinki, Gustaf Hällströmin katu 2a, FI-00560 Helsinki, Finland.

出版信息

Radiat Prot Dosimetry. 2013 Dec;157(4):515-24. doi: 10.1093/rpd/nct162. Epub 2013 Jul 3.

Abstract

The objective of this study was to assess and compare the organ and effective doses in the knee area resulting from different commercially available multislice computed tomography devices (MSCT), one cone beam computed tomography device (CBCT) and one conventional X-ray radiography device using MOSFET dosemeters and an anthropomorphic RANDO knee phantom. Measurements of the MSCT devices resulted in effective doses ranging between 27 and 48 µSv. The CBCT measurements resulted in an effective dose of 12.6 µSv. The effective doses attained using the conventional radiography device were 1.8 µSv for lateral and 1.2 µSv for anterior-posterior projections. The effective dose resulting from conventional radiography was considerably lower than those recorded for the CBCT and MSCT devices. The MSCT effective dose results were two to four times higher than those measured on the CBCT device. This study demonstrates that CBCT can be regarded as a potential low-dose 3D imaging technique for knee examinations.

摘要

本研究的目的是使用MOSFET剂量仪和拟人化的RANDO膝关节模型,评估并比较不同市售多层螺旋计算机断层扫描设备(MSCT)、一台锥形束计算机断层扫描设备(CBCT)和一台传统X射线摄影设备在膝关节区域产生的器官剂量和有效剂量。MSCT设备的测量结果显示有效剂量在27至48微希沃特之间。CBCT测量得出的有效剂量为12.6微希沃特。使用传统摄影设备获得的有效剂量,侧位投照为1.8微希沃特,前后位投照为1.2微希沃特。传统摄影产生的有效剂量显著低于CBCT和MSCT设备记录的剂量。MSCT的有效剂量结果比CBCT设备测量结果高两到四倍。本研究表明,CBCT可被视为一种用于膝关节检查的潜在低剂量三维成像技术。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验