MR-guided HIFU treatment of symptomatic uterine fibroids using novel feedback-regulated volumetric ablation: effectiveness and clinical practice.

作者信息

Ruhnke H, Eckey T, Bohlmann M K, Beldoch M P, Neumann A, Agic A, Hägele J, Diedrich K, Barkhausen J, Hunold P

机构信息

Clinic for Radiology and Nuclear Medicine, University Hospital Schleswig-Holstein, Campus Lübeck, Lübeck/Germany.

出版信息

Rofo. 2013 Sep;184(10):983-91. doi: 10.1055/s-0033-1335289. Epub 2013 Jul 24.

Abstract

PURPOSE

To evaluate a novel feedback-regulated volumetric sonication method in MR-guided HIFU treatment of symptomatic uterine fibroids.

MATERIALS AND METHODS

27 fibroids with an average volume of 124.9 ± 139.8 cc in 18 women with symptomatic uterine fibroids were ablated using the new HIFU system Sonalleve (1.5 T MR system Achieva, Philips). 21 myomas in 13 women were reevaluated 6 months later. Standard (treatment) cells (TC) and feedback-regulated (feedback) cells (FC) with a diameter of 4, 8, 12, and 16 mm were used and compared concerning sonication success, diameter of induced necrosis, and maximum achieved temperature. The non-perfused volume ratio (NPV related to myoma volume) was quantified. The fibroid volume was measured before, 1 month, and 6 months after therapy. Symptoms were quantified using a specific questionnaire (UFS-QoL).

RESULTS

In total, 205 TC and 227 FC were applied. The NPV ratio was 23 ± 15 % (2 - 55). The TC were slightly smaller than intended (-3.9 ± 52 %; range, -100 - 81), while the FC were 20.1 ± 25.3 % bigger (p = 0.02). Feedback mechanism is less diversifying in diameter (p < 0.001). Overall, the FC correlate well with the planned treatment diameter (r = 0.79), other than the TC (r = 0.38). Six months after therapy, the fibroid volume was reduced by 45 ± 21 % (5 - 100) (p = 0.001). The symptoms decreased significantly (p = 0.001). No serious adverse events were recorded.

CONCLUSION

Use of volumetric sonication leads to homogenous heating and sufficient necrosis. It is a safe and effective therapy for treating symptomatic uterine fibroids. Successful sonication of feedback cells leads to more contiguous necrosis in diameter and a less diversifying temperature.

KEY POINTS

▶ MR-guided HIFU ablation of symptomatic uterine fibroids is a valuable treatment option. ▶ By non-invasive HIFU fibroid volumes can be reduced and symptoms improved. ▶ The novel feedback-regulated treatment cells offer advantages over standard treatment cells.

摘要

目的

评估一种新型反馈调节容积式超声聚焦方法在磁共振引导高强度聚焦超声(HIFU)治疗有症状子宫肌瘤中的应用。

材料与方法

使用新型HIFU系统Sonalleve(飞利浦Achieva 1.5T磁共振系统)对18例有症状子宫肌瘤患者的27个平均体积为124.9±139.8立方厘米的肌瘤进行消融。13例患者中的21个肌瘤在6个月后进行重新评估。使用直径为4、8、12和16毫米的标准(治疗)单元(TC)和反馈调节(反馈)单元(FC),比较超声聚焦成功情况、诱导坏死直径和达到的最高温度。对无灌注体积比(与肌瘤体积相关的NPV)进行量化。在治疗前、治疗后1个月和6个月测量肌瘤体积。使用特定问卷(UFS-QoL)对症状进行量化。

结果

共应用了205个TC和227个FC。NPV比率为23±15%(2 - 55)。TC比预期略小(-3.9±52%;范围,-100 - 81),而FC大20.1±25.3%(p = 0.02)。反馈机制在直径上的差异较小(p < 0.001)。总体而言,FC与计划治疗直径相关性良好(r = 0.79),而TC则不然(r = 0.38)。治疗6个月后,肌瘤体积减少了45±21%(5 - 100)(p = 0.001)。症状显著减轻(p = 0.001)。未记录到严重不良事件。

结论

容积式超声聚焦的应用导致均匀加热和充分坏死。它是治疗有症状子宫肌瘤的一种安全有效的疗法。反馈单元的超声聚焦成功导致直径上更连续的坏死和温度差异更小。

要点

▶ 磁共振引导高强度聚焦超声消融有症状子宫肌瘤是一种有价值的治疗选择。▶ 通过非侵入性高强度聚焦超声可减少肌瘤体积并改善症状。▶ 新型反馈调节治疗单元比标准治疗单元具有优势。