Ito Yosuke, Ohnishi Hiroyuki, Kamada Keigo, Kobayashi Tetsuo
Annu Int Conf IEEE Eng Med Biol Soc. 2013;2013:3254-7. doi: 10.1109/EMBC.2013.6610235.
The goal of this study is to determine optimal conditions of optically pumped atomic magnetometers using a hybrid cell of potassium and rubidium atoms. We theoretically investigated the properties of the magnetometers and considered the optimal density ratio of potassium and rubidium atoms. We found that the experimental results agreed well with the theoretical estimations and the density ratio of potassium and rubidium atoms should be around 200 to achieve the highest sensitivity.
本研究的目标是确定使用钾原子和铷原子混合池的光泵原子磁力仪的最佳条件。我们从理论上研究了磁力仪的特性,并考虑了钾原子和铷原子的最佳密度比。我们发现实验结果与理论估计吻合良好,钾原子和铷原子的密度比应为200左右才能实现最高灵敏度。