Determination of the response function for two personal neutron dosemeter designs based on PADC.

作者信息

Mayer S, Assenmacher F, Boschung M

机构信息

Division for Radiation Safety and Security, Paul Scherrer Institute, CH-5232 Villigen PSI, Switzerland

Division for Radiation Safety and Security, Paul Scherrer Institute, CH-5232 Villigen PSI, Switzerland.

出版信息

Radiat Prot Dosimetry. 2014 Oct;161(1-4):82-5. doi: 10.1093/rpd/nct256. Epub 2013 Oct 31.

Abstract

Since 1998 neutron dosimetry based on PADC (poly allyl diglycol carbonate) is done with a so-called original Paul Scherrer Institute (PSI) design at PSI. The original design (i.e. holder) was later changed. Both designs are optimised for use in workplaces around high-energy accelerators, where the neutron energy spectra are dominated by fast neutrons ranging up to some 100 MeV. In addition to the change of the dosemeter design a new evaluation method based on a microscope scanning technique has been introduced and the etching conditions have been optimised. In the present work, the responses obtained with the original and the new dosemeter designs are compared for fields of radionuclide sources and monoenergetic reference fields using the new evaluation method. The response curves in terms of the personal dose equivalent for normally incident neutrons were built as functions of the incident neutron energy.