Institute for Environmental Chemistry, National Research Council of Canada, K1A 0R9, Ottawa, Ontario, Canada.
J Am Soc Mass Spectrom. 1992 May;3(4):281-8. doi: 10.1016/1044-0305(92)87055-4.
Aqueous metal salt solutions were used as models to probe the origin of the species observed in the electrospray mass spectrum. A qualitative or semiquantitative correlation among different species was observed between electrospray responses and calculated equilibrium aqueous solution concentrations. Quantitative correlations were obtained, however, when ions that were identical in charge and similar in type were selected for comparison. In these experiments the ions experienced very similar electrospray-related processes and their effects on the responses were canceled in a comparison of these ions. Consequently, the relative abundances of these ions in the electrospray mass spectrum closely matched the calculated relative abundances in aqueous solution. Our results suggest that the basic principle that determines ionic distribution in the electro spray mass spectrum in aqueous solution chemistry.
水相金属盐溶液被用作模型,以探究电喷雾质谱中观察到的物种的起源。在电喷雾响应和计算的平衡水相浓度之间,观察到不同物种之间存在定性或半定量的相关性。然而,当选择带相同电荷且类型相似的离子进行比较时,获得了定量相关性。在这些实验中,离子经历了非常相似的与电喷雾相关的过程,并且它们对响应的影响在这些离子的比较中被抵消了。因此,这些离子在电喷雾质谱中的相对丰度与在水溶液中的计算相对丰度非常吻合。我们的结果表明,决定水溶液中电喷雾质谱中离子分布的基本原理与水溶液化学中的相同。