Improvement of the detection of human pulpal blood flow using a laser Doppler flowmeter modified for low flow velocity.

作者信息

Qu Xiaofu, Ikawa Motohide, Shimauchi Hidetoshi

机构信息

Division of Periodontology and Endodontology, Department of Oral Biology, Tohoku University Graduate School of Dentistry, 4-1 Seiryo-machi, Aoba-ku, Sendai 980-8575, Japan; Dalian Stomatological Hospital, Changjiang Road 935, Shahekou District, Dalian 116021, China.

Division of Periodontology and Endodontology, Department of Oral Biology, Tohoku University Graduate School of Dentistry, 4-1 Seiryo-machi, Aoba-ku, Sendai 980-8575, Japan.

出版信息

Arch Oral Biol. 2014 Feb;59(2):199-206. doi: 10.1016/j.archoralbio.2013.11.009. Epub 2013 Nov 26.

Abstract

OBJECTIVE

Human pulpal blood flow (PBF) signals as measured by laser Doppler flowmeter (LDF) decrease with age. Although this decrease is considered to be due in part to slow blood flow, information regarding this velocity in humans has been lacking. The aims of the present study were to estimate the blood flow velocity in human dental pulp and to evaluate the validity of LDF modified for the measurement of slow blood flow.

DESIGN

Mean blood flow velocities at the upper central incisor, gingiva, fingertip and forearm of 28 volunteers (mean age: 38.6 years old) were estimated using LDF with a frequency analyser. Blood flow signals at these measurement areas were recorded using two different LDFs: (a) one with a standard blood flow range; and (b) one modified for low blood flow velocity.

RESULTS

The frequency range of the Doppler shift measured at the teeth with an opaque rubber dam was the narrowest (median: 4.3kHz) among all of the measurement areas. The estimated mean blood flow velocity was the slowest at the teeth with a dam (median: 0.18mm/s). LDF for low blood flow velocity detected larger and clearer pulsatile blood flow signals from the teeth with dams than did standard LDF.

CONCLUSIONS

The present results indicate that the velocity of PBF in humans is very low and that LDF modified for the measurement of slow blood flow is appropriate for PBF measurement in humans.

摘要

目的

用激光多普勒血流仪(LDF)测量的人类牙髓血流(PBF)信号会随着年龄增长而降低。尽管这种降低部分被认为是由于血流缓慢所致,但一直缺乏有关人类牙髓血流速度的信息。本研究的目的是估计人类牙髓中的血流速度,并评估为测量缓慢血流而改进的LDF的有效性。

设计

使用带有频率分析仪的LDF估计28名志愿者(平均年龄:38.6岁)上颌中切牙、牙龈、指尖和前臂的平均血流速度。在这些测量区域的血流信号使用两种不同的LDF进行记录:(a)一种具有标准血流范围;(b)一种针对低血流速度进行了改进。

结果

在所有测量区域中,用不透明橡胶障覆盖牙齿时测得的多普勒频移频率范围最窄(中位数:4.3kHz)。在有橡胶障的牙齿处估计的平均血流速度最慢(中位数:0.18mm/s)。与标准LDF相比,用于低血流速度的LDF能检测到来自有橡胶障牙齿的更大、更清晰的搏动性血流信号。

结论

目前的结果表明,人类牙髓血流速度非常低,并且为测量缓慢血流而改进的LDF适用于人类牙髓血流测量。