Leksell伽玛刀®Perfexion™患者定位系统(PPS)的长期稳定性

Long-term stability of the Leksell Gamma Knife® Perfexion™ patient positioning system (PPS).

作者信息

Novotny J, Bhatnagar J P, Xu Y, Huq M S

机构信息

Department of Radiation Oncology, Medical Physics Division, University of Pittsburgh Cancer Institute, Pittsburgh, Pennsylvania 15213; Department of Medical Physics, Na Homolce Hospital, Prague 150 30, Czech Republic; Institute of Biophysics and Informatics, First Faculty of Medicine, Charles University in Prague, Prague 120 00, Czech Republic; Department of Dosimetry and Application of Ionizing Radiation, Faculty of Nuclear Sciences and Physical Engineering, Czech Technical University in Prague, Prague 115 19, Czech Republic; and Department of Radiation Oncology, University Hospital Motol, Prague 150 06, Czech Republic.

Department of Radiation Oncology, Medical Physics Division, University of Pittsburgh Cancer Institute, Pittsburgh, Pennsylvania 15213.

出版信息

Med Phys. 2014 Mar;41(3):031711. doi: 10.1118/1.4866225.

Abstract

PURPOSE

To assess the long-term mechanical stability and accuracy of the patient positioning system (PPS) of the Leksell Gamma Knife(®) Perfexion™ (LGK PFX).

METHODS

The mechanical stability of the PPS of the LGK PFX was evaluated using measurements obtained between September 2007 and June 2011. Three methods were employed to measure the deviation of the coincidence of the radiological focus point (RFP) and the PPS calibration center point (CCP). In the first method, the onsite diode test tool with single diode detector was used together with the 4 mm collimator on a daily basis. In the second method, a service diode test tool with three diode detectors was used biannually at the time of the routine preventive maintenance. The test performed with the service diode test tool measured the deviations for all three collimators 4, 8, and 16 mm and also for three different positions of the PPS. The third method employed the conventional film pin-prick method. This test was performed annually for the 4 mm collimator at the time of the routine annual QA. To estimate the effect of the patient weight on the performance of the PPS, the focus precision tests were also conducted with varying weights on the PPS using a set of lead bricks.

RESULTS

The average deviations measured from the 641 daily focus precision tests were 0.1 ± 0.1, 0.0 ± 0.0, and 0.0 ± 0.0 mm, respectively, for the 4 mm collimator in the X (left/right of the patient), Y (anterior/posterior of the patient), and Z (superior/inferior of the patient) directions. The average of the total radial deviations as measured during ten semiannual measurements with the service diode test tool were 0.070 ± 0.029, 0.060 ± 0.022, and 0.103 ± 0.028 mm, respectively for the central, long, and short diodes for the 4 mm collimator. Similarly, the average total radial deviations measured during the semiannual measurements for the 4, 8, and 16 mm collimators and using the central diode were 0.070 ± 0.029, 0.097 ± 0.025, 0.159 ± 0.028 mm, respectively. The average values of the deviations as obtained from the five annual film pin-prick tests for the 4 mm collimator were 0.10 ± 0.06, 0.06 ± 0.09, and 0.03 ± 0.03 mm for the X, Y, Z stereotactic directions, respectively. Only a minor change was observed in the total radial deviations of the PPS as a function of the simulated patient weight up to 202 kg on the PPS.

CONCLUSIONS

Excellent long-term mechanical stability and high accuracy was observed for the PPS of the LGK PFX. No PPS recalibration or any adjustment in the PPS was needed during the monitored period of time. Similarly, the weight on the PPS did not cause any significant disturbance in the performance of the PPS for up to 202 kg simulated patient weight.

摘要

目的

评估Leksell伽玛刀Perfexion™(LGK PFX)患者定位系统(PPS)的长期机械稳定性和准确性。

方法

利用2007年9月至2011年6月期间获得的测量数据评估LGK PFX的PPS的机械稳定性。采用三种方法测量放射焦点(RFP)与PPS校准中心点(CCP)重合的偏差。第一种方法是每天使用带有单个二极管探测器的现场二极管测试工具和4毫米准直器。第二种方法是在例行预防性维护时每半年使用一次带有三个二极管探测器的服务二极管测试工具。使用服务二极管测试工具进行的测试测量了4、8和16毫米所有三个准直器以及PPS三个不同位置的偏差。第三种方法采用传统的胶片针刺法。在例行年度质量保证时每年对4毫米准直器进行此测试。为了评估患者体重对PPS性能的影响,还使用一组铅砖在PPS上施加不同重量进行焦点精度测试。

结果

在641次每日焦点精度测试中,4毫米准直器在X方向(患者左右)、Y方向(患者前后)和Z方向(患者上下)测量的平均偏差分别为0.1±0.1、0.0±0.0和0.0±0.0毫米。使用服务二极管测试工具进行的十次半年期测量中,4毫米准直器的中央、长和短二极管测量的总径向平均偏差分别为0.070±0.029、0.060±0.022和0.103±0.028毫米。同样,在4、8和16毫米准直器的半年期测量中使用中央二极管测量的总径向平均偏差分别为0.070±0.029、0.097±0.025、0.159±0.028毫米。从4毫米准直器的五次年度胶片针刺测试获得的偏差平均值在X、Y、Z立体定向方向分别为0.10±0.06、0.06±0.09和0.03±0.03毫米。在PPS上模拟患者体重高达202千克时,仅观察到PPS的总径向偏差随模拟患者体重有微小变化。

结论

观察到LGK PFX的PPS具有出色的长期机械稳定性和高精度。在监测期间无需对PPS进行重新校准或任何调整。同样,在模拟患者体重高达202千克时,PPS上的重量对PPS的性能未造成任何显著干扰。