Dose Reduction in Fluoroscopic Interventions Using a Combination of a Region of Interest (ROI) X-Ray Attenuator and Spatially-Different, Temporally-Variable Temporal Filtering.

作者信息

Vasan S N Swetadri, Pope Liza, Ionita Ciprian N, Titus A H, Bednarek D R, Rudin S

机构信息

Department of Electrical Engineering, University at Buffalo ; Toshiba Stroke and Vascular Research Center, University at Buffalo.

Toshiba Stroke and Vascular Research Center, University at Buffalo.

出版信息

Proc SPIE Int Soc Opt Eng. 2013 Mar 6;8668:86683Y. doi: 10.1117/12.2006277.

Abstract

A novel dose reduction technique for fluoroscopic interventions involving a combination of a material x-ray region of interest (ROI) attenuator and spatially different, temporally variable ROI temporal recursive filter, was used to guide the catheter to the ROI in three live animal studies, two involving rabbits and one involving a sheep. In the two rabbit studies presented, a catheter was guided to the entrance of the carotid artery. With the added ROI attenuator the image under the high attenuation region is very noisy. By using temporal filtering with a filter weight of 0.6 on previous frames, the noise is reduced. In the sheep study the catheter was guided to the descending aorta of the animal. The sheep offered a relatively higher attenuation to the incident x-rays and thus a higher temporal filter weight of 0.8 on previous frames was used during the procedure to reduce the noise to levels acceptable by the interventionalist. The image sequences from both studies show that significant dose reduction of 5-6 times can be achieved with acceptable image quality outside the ROI by using the above mentioned technique. Even though the temporal filter weighting outside the ROI is higher, the consequent lag does not prevent perception of catheter movement.

摘要

相似文献

4
5
Scatter reduction for high resolution image detectors with a region of interest attenuator.
Proc SPIE Int Soc Opt Eng. 2014 Mar 19;9033:903364. doi: 10.1117/12.2043061.
6
Moving segments region of interest attenuator for x-ray fluoroscopy.
Med Phys. 1999 Jan;26(1):19-26. doi: 10.1118/1.598474.
7
A Patient Dose-Reduction Technique for Neuroendovascular Image-Guided Interventions: Image-Quality Comparison Study.
AJNR Am J Neuroradiol. 2018 Apr;39(4):734-741. doi: 10.3174/ajnr.A5552. Epub 2018 Feb 15.
8
Dose reduction by moving a region of interest (ROI) beam attenuator to follow a moving object of interest.
Proc SPIE Int Soc Opt Eng. 2012;8313:831355. doi: 10.1117/12.911136. Epub 2012 Feb 23.
10
Organ and effective dose reduction for region-of-interest (ROI) CBCT and fluoroscopy.
Proc SPIE Int Soc Opt Eng. 2017 Feb 11;10132. doi: 10.1117/12.2254061. Epub 2017 Mar 9.

本文引用的文献

2
Endovascular image-guided interventions (EIGIs).
Med Phys. 2008 Jan;35(1):301-9. doi: 10.1118/1.2821702.
3
Optimization of variable temporal averaging in digital fluoroscopy.
Br J Radiol. 2004 Aug;77(920):675-8. doi: 10.1259/bjr/72726487.
4
Automatic filter placement device for region of interest (ROI) fluoroscopy.
Health Phys. 1997 Jan;72(1):141-6. doi: 10.1097/00004032-199701000-00020.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验