Transit dosimetry in dynamic IMRT with an a-Si EPID.

作者信息

Sabet Mahsheed, Rowshanfarzad Pejman, Menk Fred W, Greer Peter B

机构信息

Department of Radiation Oncology, Sir Charles Gairdner Hospital, Hospital Avenue, Nedlands, WA, 6009, Australia,

出版信息

Med Biol Eng Comput. 2014 Jul;52(7):579-88. doi: 10.1007/s11517-014-1161-y. Epub 2014 May 31.

Abstract

Using an amorphous silicon (a-Si) EPID for transit dosimetry requires detailed characterization of its dosimetric response in a variety of conditions. In this study, a measurement-based model was developed to calibrate an a-Si EPID response to dose for transit dosimetry by comparison with a reference ionization chamber. The ionization chamber reference depth and the required additional buildup thickness for electronic portal imaging devices (EPID) transit dosimetry were determined. The combined effects of changes in radiation field size, phantom thickness, and the off-axis distance on EPID transit dosimetry were characterized. The effect of scattered radiation on out-of-field response was investigated for different field sizes and phantom thicknesses by evaluation of the differences in image profiles and in-water measured profiles. An algorithm was developed to automatically apply these corrections to EPID images based on the user-specified field size and phantom thickness. The average phantom thickness and an effective field size were used for IMRT fields, and images were acquired in cine mode in the presence of an anthropomorphic phantom. The effective field size was defined as the percentage of the jaw-defined field that was involved during the delivery. Nine head and neck dynamic IMRT fields were tested by comparison with a MatriXX two-dimensional array dosimeter using the Gamma (3%, 3 mm) evaluation. A depth of 1.5 cm was selected as the ionization chamber reference depth. An additional 2.2 mm of copper buildup was added to the EPID. Comparison of EPID and MatriXX dose images for the tested fields showed that using a 10% threshold, the average number of points with Gamma index <1 was 96.5%. The agreement in the out-of field area was shown by selection of a 2% threshold which on average resulted in 94.8% of points with a Gamma index <1. The suggested method is less complicated than previously reported techniques and can be used for all a-Si EPIDs regardless of the manufacturer.

摘要

使用非晶硅(a-Si)电子射野影像装置(EPID)进行传输剂量测定需要详细表征其在各种条件下的剂量响应。在本研究中,通过与参考电离室比较,开发了一种基于测量的模型,用于校准a-Si EPID对传输剂量测定剂量的响应。确定了电离室参考深度以及电子射野影像装置(EPID)传输剂量测定所需的额外建成厚度。表征了辐射野大小、模体厚度和离轴距离变化对EPID传输剂量测定的综合影响。通过评估图像轮廓和水中测量轮廓的差异,研究了不同野大小和模体厚度下散射辐射对野外响应的影响。开发了一种算法,可根据用户指定的野大小和模体厚度自动对EPID图像应用这些校正。对于调强放疗(IMRT)野,使用平均模体厚度和有效野大小,并在存在人体模体的情况下以电影模式采集图像。有效野大小定义为在照射过程中涉及的限束器定义野的百分比。通过使用Gamma(3%,3 mm)评估与MatriXX二维阵列剂量仪比较,测试了九个头颈部动态IMRT野。选择1.5 cm深度作为电离室参考深度。向EPID添加了额外2.2 mm的铜建成。测试野的EPID和MatriXX剂量图像比较表明,使用10%阈值时,Gamma指数<1的点的平均数量为96.5%。通过选择2%阈值显示了野外区域的一致性,平均导致94.8%的点Gamma指数<1。所建议的方法比先前报道的技术更简单,并且可用于所有a-Si EPID,无论制造商如何。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验