Suppr超能文献

Non-contact direct measurement of the magnetocaloric effect in thin samples.

作者信息

Cugini F, Porcari G, Solzi M

机构信息

Department of Physics and Earth Sciences, University of Parma, Parco Area delle Scienze 7/A, 43124 Parma, Italy.

出版信息

Rev Sci Instrum. 2014 Jul;85(7):074902. doi: 10.1063/1.4890394.

Abstract

An experimental setup, based on a non-contact temperature sensor, is proposed to directly measure the magnetocaloric effect of samples few micrometers thick. The measurement of the adiabatic temperature change of foils and ribbons is fundamental to design innovative devices based on magnetocaloric thin materials or micro-structuring bulk samples. The reliability of the proposed setup is demonstrated by comparing the measurements performed on a bulk gadolinium sample with the results obtained by an experimental setup based on a Cernox bare chip thermoresistance and by in-field differential scanning calorimetry. We show that this technique can measure the adiabatic temperature variation on gadolinium sheets as thin as 27 μm. Heat transfer simulations are added to describe the capability of the presented technique.