Akino Yuichi, Gibbons John P, Neck Daniel W, Chu Connel, Das Indra J
Indiana University School of Medicine.
J Appl Clin Med Phys. 2014 Jul 8;15(4):4850. doi: 10.1120/jacmp.v15i4.4850.
Accurate beam data acquisition during commissioning is essential for modeling the treatment planning system and dose calculation in radiotherapy. Although currently several commercial scanning systems are available, there is no report that compared the differences among the systems because most institutions do not acquire several scanning systems due to the high cost, storage space, and infrequent usage. In this report, we demonstrate the intra- and intervariability of beam profiles measured with four commercial scanning systems. During a recent educational and training workshop, four different vendors of beam scanning water phantoms were invited to demonstrate the operation and data collection of their systems. Systems were set up utilizing vendor-recommended protocols and were operated with a senior physicist, who was assigned as an instructor along with vendor. During the training sessions, each group was asked to measure beam parameters, and the intravariability in percent depth dose (PDD). At the end of the day, the profile of one linear accelerator was measured with each system to evaluate intervariability. Relatively very small (SD < 0.12%) intervariability in PDD was observed among four systems at a region deeper than peak (1.5 cm). All systems showed almost identical profiles. At the area within 80% of radiation field, the average, and maximum differences were within ± 0.35% and 0.80%, respectively, compared to arbitrarily chosen IBA system as reference. In the penumbrae region, the distance to agreement (DTA) of the region where dose difference exceed ± 1% was less than 1 mm. Repeated PDD measurement showed small intravariability with SD < 0.5%, although large SD was observed in the buildup region. All four water phantom scanning systems demonstrated adequate accuracy for beam data collection (i.e., within 1% of dose difference or 1 mm of DTA among each other). It is concluded that every system is capable of acquiring accurate beam. Thus the selection of a water scanning system should be based on institutional comfort, personal preference of software and hardware, and financial consideration.
在调试验收期间准确采集射束数据对于放疗治疗计划系统建模和剂量计算至关重要。尽管目前有几种商用扫描系统可供使用,但尚无报告比较这些系统之间的差异,因为由于成本高、存储空间大以及使用频率低,大多数机构不会购置多种扫描系统。在本报告中,我们展示了用四种商用扫描系统测量的射束剖面的内部和相互之间的变异性。在最近一次教育和培训研讨会上,邀请了四家不同的射束扫描水模供应商展示其系统的操作和数据收集。系统按照供应商推荐的协议进行设置,并由一名资深物理学家操作,该物理学家与供应商人员一起被指定为指导教师。在培训期间,要求每个小组测量射束参数以及百分深度剂量(PDD)的内部变异性。在当天结束时,用每个系统测量一台直线加速器的剖面以评估相互之间的变异性。在深度超过峰值(1.5厘米)的区域,四个系统之间观察到PDD的相互变异性相对非常小(标准差<0.12%)。所有系统显示出几乎相同的剖面。与任意选择的IBA系统作为参考相比,在辐射野的80%范围内,平均差异和最大差异分别在±0.35%和0.80%以内。在半值层区域,剂量差异超过±1%的区域的一致性距离(DTA)小于1毫米。重复的PDD测量显示内部变异性较小,标准差<0.5%,尽管在建成区观察到较大的标准差。所有四个水模扫描系统在射束数据采集方面都显示出足够的准确性(即彼此之间剂量差异在1%以内或DTA在1毫米以内)。得出的结论是每个系统都能够采集准确的射束。因此,水扫描系统的选择应基于机构的便利性、对软件和硬件的个人偏好以及财务考虑。