Butler G C, Al-Assaf N, Tarrant A, Ryan S, El-Khuffash A
Ir Med J. 2014 Sep;107(8):256-8.
We aimed to assess the difference in measurement of the distance of the UVC tip from the diaphragm between (Anteroposterior) AP and lateral radiographs and to determine the reliability of the measurement of UVC tip distance from the diaphragm between the two views. A retrospective review of paired AP and lateral radiographs taken to assess UVC tip position was carried out in 25 infants was conducted and reliability analysis was carried out. There was a significant difference in the mean (SD) distance of the UVC catheter above the diaphragm between the AP and lateral radiographs: 8.7 (7.8) mm versus 11.6 (7.3) mm (p = 0.003) respectively. Measurements using lateral radiographs were more reliable (Intraclass correlation coefficient: 0.99 vs. 0.93). Inter-observer reliability analysis yielded similar results. Lateral radiographs are more reliable in measurement of UVC tip position and should be performed in conjunction with AP films to aid in determining UVC position.
我们旨在评估前后位(AP)X线片与侧位X线片在测量脐静脉导管(UVC)尖端至膈肌距离方面的差异,并确定两种视图下测量UVC尖端至膈肌距离的可靠性。对25例婴儿为评估UVC尖端位置而拍摄的配对AP和侧位X线片进行了回顾性分析,并进行了可靠性分析。AP和侧位X线片上UVC导管高于膈肌的平均(标准差)距离存在显著差异:分别为8.7(7.8)mm和11.6(7.3)mm(p = 0.003)。使用侧位X线片测量更可靠(组内相关系数:0.99对0.93)。观察者间可靠性分析得出了类似结果。侧位X线片在测量UVC尖端位置方面更可靠,应与AP片一起进行以辅助确定UVC位置。