Copenhagen University Hospital, Frederiksberg, Denmark, and Copenhagen University Hospital at Glostrup, Glostrup, Denmark.
Arthritis Rheumatol. 2015 Feb;67(2):386-95. doi: 10.1002/art.38940.
To determine how settings for power and color Doppler ultrasound sensitivity vary on different high- and intermediate-range ultrasound machines and to evaluate the impact of these changes on Doppler scoring of inflamed joints.
Six different types of ultrasound machines were used. On each machine, the factory setting for superficial musculoskeletal scanning was used unchanged for both color and power Doppler modalities. The settings were then adjusted for increased Doppler sensitivity, and these settings were designated study settings. Eleven patients with rheumatoid arthritis (RA) with wrist involvement were scanned on the 6 machines, each with 4 settings, generating 264 Doppler images for scoring and color quantification. Doppler sensitivity was measured with a quantitative assessment of Doppler activity: color fraction. Higher color fraction indicated higher sensitivity.
Power Doppler was more sensitive on half of the machines, whereas color Doppler was more sensitive on the other half, using both factory settings and study settings. There was an average increase in Doppler sensitivity, despite modality, of 78% when study settings were applied. Over the 6 machines, 2 Doppler modalities, and 2 settings, the grades for each of 7 of the patients varied between 0 and 3, while the grades for each of the other 4 patients varied between 0 and 2.
The effect of using different machines, Doppler modalities, and settings has a considerable influence on the quantification of inflammation by ultrasound in RA patients, and this must be taken into account in multicenter studies.
确定不同高中档超声设备的功率和彩色多普勒超声灵敏度设置如何变化,并评估这些变化对炎症关节多普勒评分的影响。
使用了 6 种不同类型的超声机。在每台机器上,均使用不变的浅表肌肉骨骼扫描工厂设置,用于彩色和功率多普勒模式。然后调整设置以增加多普勒灵敏度,这些设置被指定为研究设置。对 11 例腕部受累的类风湿关节炎(RA)患者在 6 台机器上进行扫描,每台机器有 4 种设置,共产生 264 个用于评分和彩色量化的多普勒图像。使用多普勒活动的定量评估:彩色分数来测量多普勒灵敏度。较高的彩色分数表示较高的灵敏度。
使用工厂设置和研究设置时,有一半的机器功率多普勒更敏感,而另一半机器彩色多普勒更敏感。尽管采用了不同的模式和设置,多普勒灵敏度平均增加了 78%。在 6 台机器、2 种多普勒模式和 2 种设置中,7 名患者中的每一名的分数在 0 到 3 之间变化,而另外 4 名患者中的每一名的分数在 0 到 2 之间变化。
使用不同的机器、多普勒模式和设置对 RA 患者的超声炎症定量有很大影响,在多中心研究中必须考虑到这一点。