[彩色多普勒超声体外测量血管内血流]

[In vitro measurement of intravascular blood flow with color Doppler sonography].

作者信息

Landwehr P, Dölken W, Lackner K

机构信息

Institut für Röntgendiagnostik der Universität Würzburg.

出版信息

Rofo. 1989 Feb;150(2):192-7. doi: 10.1055/s-2008-1047004.

Abstract

The accuracy of colour Doppler sonography for measuring various aspects of intravascular flow (volume and velocity), under varying conditions, was studied experimentally. The effect of variable Doppler angles, different flow paths, frequency and vessel diameter on flow profiles was examined. At all levels of measurement (range 14 ml/min to 566 ml/min) there was a very close correlation between the actual and measured values (r greater than 0.99) and high accuracy. In the detection of very slow flow rates, the use of a specific high pass filter technique was able to quantify blood flow velocities of just under 1 cm./second. Multigated colour Doppler sonography is better for measuring flow volumes than a single channel duplex system.

摘要

通过实验研究了彩色多普勒超声在不同条件下测量血管内血流各方面(流量和速度)的准确性。研究了可变多普勒角度、不同血流路径、频率和血管直径对血流剖面的影响。在所有测量水平(范围为14毫升/分钟至566毫升/分钟),实际值与测量值之间存在非常紧密的相关性(r大于0.99)且准确性很高。在检测非常缓慢的流速时,使用特定的高通滤波技术能够量化略低于1厘米/秒的血流速度。多门控彩色多普勒超声在测量血流量方面比单通道双功系统更好。