Mazaheri Y, Afaq A A, Jung S I, Goldman D A, Wang L, Aslan H, Zelefsky M J, Akin O, Hricak H
Department of Medical Physics, Memorial Sloan-Kettering Cancer Center, New York, NY, USA; Department of Radiology, Memorial Sloan-Kettering Cancer Center, New York, NY, USA.
Department of Radiology, Memorial Sloan-Kettering Cancer Center, New York, NY, USA.
Clin Radiol. 2015 Apr;70(4):379-86. doi: 10.1016/j.crad.2014.12.002. Epub 2014 Dec 29.
To compare prostate volumes and distances between anatomical landmarks on MRI images obtained with a phased-array coil (PAC) only and with a PAC and an endorectal coil (ERC).
Informed consent was waived for this Health Insurance Portability and Accountability Act-compliant study. Fifty-nine men underwent PAC-MRI and ERC-MRI at 1.5 (n = 3) or 3 T (n = 56). On MRI images, two radiologists independently measured prostate volume and distances between the anterior rectal wall (ARW) and symphysis pubis at the level of the verumontanum; ARW and symphysis pubis at the level of the mid-symphysis pubis; and bladder neck and mid-symphysis pubis. Differences between measurements from PAC-MRI and ERC-MRI were assessed with the Wilcoxon RANK SUM test. Inter-reader agreement was assessed using the concordance correlation coefficient (CCC).
Differences in prostate volume between PAC-MRI and ERC-MRI [median: -0.75 mm(3) (p = 0.10) and median: -0.84 mm(3) (p = 0.06) for readers 1 and 2, respectively] were not significant. For readers 1 and 2, median differences between distances were as follows: -10.20 and -12.75 mm, respectively, ARW to symphysis pubis at the level of the verumontanum; -6.60 and -6.08 mm, respectively, ARW to symphysis pubis at the level of the mid-symphysis pubis; -3 and -3 mm respectively, bladder neck to mid-symphysis pubis. All differences in distance were significant for both readers (p ≤ 0.0005). Distances were larger on PAC-MRI (p ≤ 0.0005). Inter-reader agreement regarding prostate volume was almost perfect on PAC-MRI (CCC: 0.99; 95% CI: 0.98-1.00) and ERC-MRI (CCC: 0.99; 95% CI: 0.99-1.00); inter-reader agreement for distance measurements varied (CCCs: 0.54-0.86).
Measurements of distances between anatomical landmarks differed significantly between ERC-MRI and PAC-MRI, although prostate volume measurements did not.
比较仅使用相控阵线圈(PAC)以及使用PAC和直肠内线圈(ERC)获取的MRI图像上的前列腺体积和解剖标志之间的距离。
对于这项符合《健康保险流通与责任法案》的研究,无需获取知情同意书。59名男性接受了1.5T(n = 3)或3T(n = 56)的PAC-MRI和ERC-MRI检查。在MRI图像上,两名放射科医生独立测量前列腺体积以及精阜水平处直肠前壁(ARW)与耻骨联合之间的距离;耻骨联合中点水平处的ARW与耻骨联合之间的距离;以及膀胱颈与耻骨联合中点之间的距离。使用Wilcoxon秩和检验评估PAC-MRI和ERC-MRI测量结果之间的差异。使用一致性相关系数(CCC)评估阅片者间的一致性。
PAC-MRI和ERC-MRI之间前列腺体积的差异[读者1的中位数:-0.75 mm³(p = 0.10),读者2的中位数:-0.84 mm³(p = 0.06)]不显著。对于读者1和读者2,距离的中位数差异如下:精阜水平处ARW至耻骨联合分别为-10.20和-12.75 mm;耻骨联合中点水平处ARW至耻骨联合分别为-6.60和-6.08 mm;膀胱颈至耻骨联合中点分别为-3和-3 mm。两位读者的所有距离差异均显著(p≤0.0005)。PAC-MRI上的距离更大(p≤0.0005)。阅片者间关于前列腺体积的一致性在PAC-MRI(CCC:0.99;95%CI:0.98 - 1.00)和ERC-MRI(CCC:0.99;95%CI:0.99 - 1.00)上几乎完美;距离测量的阅片者间一致性有所不同(CCC:范围0.54 - 0.86)。
尽管前列腺体积测量无差异,但ERC-MRI和PAC-MRI之间解剖标志之间距离的测量存在显著差异。