Research on temperature measurement technology for graphite-cone-absorption-cavity absolute calorimeter.

作者信息

Wei Ji Feng, Lu Fei, Sun Li Qun, Zhang Kai, Hu Xiao Yang, Zhou Shan, Xu De

机构信息

State Key Laboratory of Precision Measurement Technology and Instrument, Tsinghua University, Beijing 100084, People's Republic of China.

Institute of Applied Electronics, China Academy of Engineering Physics, Mianyang 621900, People's Republic of China.

出版信息

Rev Sci Instrum. 2015 Feb;86(2):025001. doi: 10.1063/1.4906934.

Abstract

The nonlinear effect of materials and sensors in high-energy laser calorimeters is especially obvious-due to the steep temperature gradients of their absorbers. Significant measurement errors occur when traditional integral temperature sensors and methods are utilized. In an effort to remedy this, a method is proposed in this paper in which an absorption cavity is divided into many parts and multiple discrete thermocouple sensors are used to measure the temperature rise of the absorbers. The temperature distribution in the absorbers is theoretically analyzed, numerically simulated, and verified through experimentation. Energy measurement results are compared according to the temperature distribution for different layouts of thermocouples. A high-accuracy calorimeter is developed by setting and optimizing thermocouple layout, as well as correcting various elements such as the specific heat of graphite and responsivity of thermocouples. The calorimeter employing this measurement method is calibrated against a standard energy meter, resulting in correction coefficient of 1.027 and relative standard deviation of the correction coefficient of only 0.8%. Theoretical analysis, numerical simulation, and experimental verification all prove that the proposed method successfully improves measurement accuracy.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验