截肢者专用袜子:袜子厚度会随正常使用而变化。

Amputee socks: Sock thickness changes with normal use.

作者信息

Cagle John C, D'Silva Krittika J, Hafner Brian J, Harrison Daniel S, Sanders Joan E

机构信息

University of Washington, Seattle, WA, USA.

University of Washington, Seattle, WA, USA

出版信息

Prosthet Orthot Int. 2016 Jun;40(3):329-35. doi: 10.1177/0309364614568412. Epub 2015 Mar 2.

Abstract

BACKGROUND

Prosthetic socks are expected to decrease in thickness and have reduced volume accommodation with normal use. It is unknown, however, to what degree they reduce in thickness over time.

OBJECTIVE

The goal of this study was to determine a correlation between the age of a prosthetic sock (defined as the out-of-package time) and the resulting change in thickness under standardized weight-bearing and non-weight-bearing conditions.

STUDY DESIGN

Experimental, mechanical assessment.

METHODS

Used prosthetic socks were donated by donors with transtibial amputation. Sock thickness was measured on a custom instrument under conditions representative of normal use. Stress-thickness response was compared to that of equivalent new socks to quantify the effects of use on sock performance.

RESULTS

Sock thickness changed non-linearly over time. On average, socks were 75% ± 17% of their initial thickness after 1 month, while socks older than 1 month were 72% ± 18% of their initial thickness. The elasticity of socks did not change with age.

CONCLUSION

The volume accommodation provided by used socks cannot be reliably predicted by ply or age. Direct measurement of total sock thickness may provide meaningful insight to quantify prosthetic users' socket fit and guide volume accommodation recommendations.

CLINICAL RELEVANCE

The mean difference in thickness between 3-ply and 5-ply used socks was equal to the standard deviation of each ply group (0.3mm). Therefore, it is possible that a 3-ply sock worn for as a little as 1 month could have a greater thickness than a 5-ply sock worn for a month.

摘要

背景

义肢袜在正常使用过程中预计会变薄且容纳体积减小。然而,其随时间变薄的程度尚不清楚。

目的

本研究的目的是确定义肢袜的使用时长(定义为开封时间)与在标准化负重和非负重条件下厚度变化之间的相关性。

研究设计

实验性机械评估。

方法

经胫骨截肢的捐赠者捐赠了使用过的义肢袜。在代表正常使用的条件下,使用定制仪器测量袜子厚度。将应力 - 厚度响应与同等新袜子进行比较,以量化使用对袜子性能的影响。

结果

袜子厚度随时间呈非线性变化。平均而言,使用1个月后,袜子厚度为初始厚度的75%±17%,而使用超过1个月的袜子厚度为初始厚度的72%±18%。袜子的弹性不会随使用时长而改变。

结论

使用过的袜子所提供的容纳体积无法通过层数或使用时长可靠预测。直接测量袜子总厚度可能有助于深入了解义肢使用者的接受腔适配情况,并指导关于容纳体积的建议。

临床意义

3层和5层使用过的袜子之间的平均厚度差异等于每个层数组的标准差(0.3毫米)。因此,使用仅1个月的3层袜子厚度可能比使用1个月的5层袜子更大。

相似文献

1
Amputee socks: Sock thickness changes with normal use.
Prosthet Orthot Int. 2016 Jun;40(3):329-35. doi: 10.1177/0309364614568412. Epub 2015 Mar 2.
2
Amputee socks: thickness of multiple socks.
Prosthet Orthot Int. 2014 Oct;38(5):405-12. doi: 10.1177/0309364613506915. Epub 2013 Nov 15.
3
Self-reported prosthetic sock use among persons with transtibial amputation.
Prosthet Orthot Int. 2014 Aug;38(4):321-31. doi: 10.1177/0309364613499064. Epub 2013 Aug 28.
4
Amputee socks: how does sock ply relate to sock thickness?
Prosthet Orthot Int. 2012 Mar;36(1):77-86. doi: 10.1177/0309364611431290. Epub 2012 Jan 6.
5
Device to monitor sock use in people using prosthetic limbs: technical report.
J Rehabil Res Dev. 2012;49(8):1229-38. doi: 10.1682/jrrd.2011.09.0169.
7
How do sock ply changes affect residual-limb fluid volume in people with transtibial amputation?
J Rehabil Res Dev. 2012;49(2):241-56. doi: 10.1682/jrrd.2011.02.0022.
9
Effects of activity intensity, time, and intermittent doffing on daily limb fluid volume change in people with transtibial amputation.
Prosthet Orthot Int. 2019 Feb;43(1):28-38. doi: 10.1177/0309364618785729. Epub 2018 Jul 16.
10
Feasibility testing of a novel prosthetic socket sensor system.
Disabil Rehabil. 2023 Jul;45(14):2374-2381. doi: 10.1080/09638288.2022.2093997. Epub 2022 Jul 7.

引用本文的文献

1
Assessing Socket Fit Effects on Pressure and Shear at a Transtibial Residuum/Socket Interface.
Appl Bionics Biomech. 2023 Aug 16;2023:3257059. doi: 10.1155/2023/3257059. eCollection 2023.
2
Characterization of Prosthetic Liner Products for People with Transtibial Amputation.
J Prosthet Orthot. 2018 Oct;30(4):187-199. doi: 10.1097/JPO.0000000000000205.
3
A finite element model to assess transtibial prosthetic sockets with elastomeric liners.
Med Biol Eng Comput. 2018 Jul;56(7):1227-1240. doi: 10.1007/s11517-017-1758-z. Epub 2017 Dec 13.
4
Effects of socket size on metrics of socket fit in trans-tibial prosthesis users.
Med Eng Phys. 2017 Jun;44:32-43. doi: 10.1016/j.medengphy.2017.03.003. Epub 2017 Apr 1.

本文引用的文献

1
Self-reported prosthetic sock use among persons with transtibial amputation.
Prosthet Orthot Int. 2014 Aug;38(4):321-31. doi: 10.1177/0309364613499064. Epub 2013 Aug 28.
2
Amputee socks: how does sock ply relate to sock thickness?
Prosthet Orthot Int. 2012 Mar;36(1):77-86. doi: 10.1177/0309364611431290. Epub 2012 Jan 6.
3
Prosthetic intervention effects on activity of lower-extremity amputees.
Arch Phys Med Rehabil. 2006 May;87(5):717-22. doi: 10.1016/j.apmr.2006.02.007.
4
Standing interface stresses as a predictor of walking interface stresses in the trans-tibial prosthesis.
Prosthet Orthot Int. 2001 Apr;25(1):34-40. doi: 10.1080/03093640108726566.