Abraham Christine L, Bangerter Neal K, McGavin Lance S, Peters Christopher L, Drew Alex J, Hanrahan Christopher J, Anderson Andrew E
Department of Bioengineering, University of Utah, Salt Lake City, Utah, USA.
Department of Orthopaedics, University of Utah, Salt Lake City, Utah, USA.
J Magn Reson Imaging. 2015 Nov;42(5):1329-38. doi: 10.1002/jmri.24902. Epub 2015 Apr 6.
To deploy and quantify the accuracy of 3D dual echo steady state (DESS) MR arthrography with hip traction to image acetabular cartilage. Clinical magnetic resonance imaging (MRI) sequences used to image hip cartilage often have reduced out-of-plane resolution and may lack adequate signal-to-noise to image cartilage.
Saline was injected into four cadaver hips placed under traction. 3D DESS MRI scans were obtained before and after cores of cartilage were harvested from the acetabulum; the two MRIs were spatially aligned to reference core positions. The thickness of cartilage cores was measured under microscopy to serve as the reference standard. 3D reconstructions of cartilage and subchondral bone were generated using automatic and semiautomatic image segmentation. Cartilage thickness estimated from the 3D reconstructions was compared to physical measurements using Bland-Altman plots.
As revealed by the automatic segmentation mask, saline imbibed the joint space throughout the articulating surface, with the exception of the posteroinferior region in two hips. Locations where air bubbles were introduced and regions of suspected low density bone disrupted an otherwise smooth automatic segmentation mask. Automatic and semiautomatic segmentation yielded a bias ± repeatability coefficient (95% limits of agreement) of 0.10 ± 0.51 mm (-0.41 to 0.61 mm) and 0.06 ± 0.43 mm (-0.37 to 0.49 mm), respectively.
Cartilage thickness can be estimated to within ∼0.5 mm of the physical value with 95% confidence using 3D reconstructions of 3D DESS MR arthrography images. Manual correction of the automatic segmentation mask may improve reconstruction accuracy.
应用髋关节牵引的三维双回波稳态(DESS)磁共振关节造影术对髋臼软骨进行成像并量化其准确性。用于髋关节软骨成像的临床磁共振成像(MRI)序列通常具有降低的平面外分辨率,并且可能缺乏足够的信噪比来对软骨进行成像。
向四个处于牵引状态的尸体髋关节内注入生理盐水。在从髋臼采集软骨芯之前和之后获取三维DESS MRI扫描;将这两次MRI在空间上与参考芯位置对齐。在显微镜下测量软骨芯的厚度作为参考标准。使用自动和半自动图像分割生成软骨和软骨下骨的三维重建。使用Bland-Altman图将从三维重建估计的软骨厚度与物理测量值进行比较。
如自动分割掩码所示,生理盐水在整个关节表面吸收关节间隙,但两个髋关节的后下区域除外。引入气泡的位置和疑似低密度骨区域破坏了原本平滑的自动分割掩码。自动和半自动分割产生的偏差±重复性系数(95%一致性界限)分别为0.10±0.51毫米(-0.41至0.61毫米)和0.06±0.43毫米(-0.37至0.49毫米)。
使用三维DESS磁共振关节造影图像的三维重建,可以在95%置信度下将软骨厚度估计在物理值的约0.5毫米范围内。手动校正自动分割掩码可能会提高重建准确性。