Abu-Rabie Paul, Denniff Philip, Spooner Neil, Chowdhry Babur Z, Pullen Frank S
†Bioanalytical Science and Toxicokinetics, Drug Metabolism and Pharmacokinetics, Platform Technologies and Sciences, GlaxoSmithKline Research and Development, Ware, SG12 0DP, United Kingdom.
‡Faculty of Engineering and Science, University of Greenwich, Medway Campus, Chatham Maritime, Kent ME4 4TB, United Kingdom.
Anal Chem. 2015;87(9):4996-5003. doi: 10.1021/acs.analchem.5b00908. Epub 2015 Apr 24.
Hematocrit (HCT)-based assay bias (composed of area and recovery bias) is an important contributing factor to the barriers that currently hinder the development and acceptance of dried blood spots (DBS) as a widely used quantitative bioanalytical sampling technique for regulatory studies. This article describes the evaluation of a practical internal standard spray addition technique, used prior to LC-MS/MS analysis, which is demonstrated to nullify the effect of recovery bias. To our knowledge, this is the first time a potential solution to HCT-based recovery bias has been investigated in detail and reported in the literature. This new technique is coupled with accurate volume DBS sampling, whole-spot extraction, and automated direct elution techniques to demonstrate a workflow that both nullifies HCT-based assay bias and the additional manual extraction burden associated with DBS analysis.
基于血细胞比容(HCT)的分析偏差(由面积偏差和回收率偏差组成)是当前阻碍干血斑(DBS)作为一种广泛应用于监管研究的定量生物分析采样技术发展和被接受的障碍的一个重要因素。本文描述了一种实用的内标喷雾添加技术的评估,该技术在液相色谱 - 串联质谱(LC-MS/MS)分析之前使用,已证明可消除回收率偏差的影响。据我们所知,这是首次对基于HCT的回收率偏差的潜在解决方案进行详细研究并在文献中报道。这项新技术与精确体积的DBS采样、全斑提取和自动直接洗脱技术相结合,展示了一种既能消除基于HCT的分析偏差又能消除与DBS分析相关的额外手动提取负担的工作流程。