使用自由手 SPECT 3D 映射在放射性引导手术中的训练方案可提高准确性和可重复性。

Improved Accuracy and Reproducibility Using a Training Protocol for Freehand-SPECT 3D Mapping in Radio-Guided Surgery.

机构信息

From the Department of Nuclear Medicine, The Netherlands Cancer Institute, Antoni van Leeuwenhoek Hospital, Amsterdam, The Netherlands.

出版信息

Clin Nucl Med. 2015 Sep;40(9):e457-60. doi: 10.1097/RLU.0000000000000787.

Abstract

PURPOSE

Freehand single-photon emission computed tomography (SPECT) (fh-SPECT) for optical tracked 3-dimensional radioactivity mapping raises popularity for sentinel node biopsy and other radio-guided procedures. However, the device seems to be complex, requiring both training and a standardized scan protocol. Therefore, the aim of this study was to compare handling reproducibility between novice and expert users; subsequently, using additional scans, we evaluated a standardized scan protocol to be tested in novice users after receiving training.

MATERIALS AND METHODS

Three groups (untrained novices, experts, and trained novices), each one composed of 5 fh-SPECT users, were given the same short introduction and the assignment to perform 3 fh-SPECT scans with the same time per scan. For the scans, a reproducible phantom with an iodine-125 point source was used. Furthermore, we performed probe-trajectory evaluation including pattern, speed, and length of scans based on recorded videos and digital target tracking.

RESULTS

The training period encompasses 30 minutes per novice user. The mean error in 3 directions based on 45 measurements was 7.4 mm for untrained novices and 3.2 mm for expert users. The trained novice group had a mean error of 2.9 mm (a significant 61% reduction), comparable to the expert group. The reproducibility values, expressed in SD, were 4.1 mm for the untrained novice group, 1.3 mm for the expert group, and 1.1 mm for the trained novices (73% improvement). The standardized scan protocol was demonstrated by means of trajectory evaluation, a scan path better systematized.

CONCLUSION

Scanning with fh-SPECT for radio-guided surgery requires a specific training period to increase handling and reproducibility of 3-dimensional imaging, and to enhance its accuracy in the operating room.

摘要

目的

自由手单光子发射计算机断层扫描(fh-SPECT)用于光学跟踪三维放射性测绘,提高了前哨淋巴结活检和其他放射性引导手术的普及度。然而,该设备似乎较为复杂,需要经过培训和标准化扫描方案。因此,本研究旨在比较新手和专家用户之间的操作可重复性;随后,我们使用额外的扫描来评估一个标准化的扫描方案,以便在接受培训后对新手用户进行测试。

材料和方法

三组(未经训练的新手、专家和经过训练的新手,每组由 5 名 fh-SPECT 用户组成)接受相同的简短介绍,并被分配执行 3 次 fh-SPECT 扫描,每次扫描的时间相同。对于扫描,使用具有碘-125 点源的可重复 phantom。此外,我们根据记录的视频和数字目标跟踪对探头轨迹进行评估,包括扫描模式、速度和长度。

结果

每位新手用户的培训期为 30 分钟。基于 45 次测量的 3 个方向的平均误差为未经训练的新手组 7.4 毫米,专家组为 3.2 毫米。经过培训的新手组的平均误差为 2.9 毫米(显著降低 61%),与专家组相当。以标准差表示的可重复性值,未经训练的新手组为 4.1 毫米,专家组为 1.3 毫米,经过培训的新手组为 1.1 毫米(提高 73%)。通过轨迹评估显示,标准化扫描方案的扫描路径更加系统化。

结论

fh-SPECT 用于放射性引导手术的扫描需要经过特定的培训期,以提高三维成像的操作和可重复性,并提高手术室中的准确性。