†Animal Drugs Research Center and ‡Denver Laboratory, U.S. Food and Drug Administration, Denver Federal Center Building 20, West Sixth Avenue and Kipling Street, Denver, Colorado 80225-0087, United States.
J Agric Food Chem. 2015 May 20;63(19):4790-8. doi: 10.1021/acs.jafc.5b00959. Epub 2015 May 7.
The objective of this study was to perform a preliminary investigation of the nontargeted search and quantitative capabilities of a single-stage Exactive High-Resolution Mass Spectrometer (HRMS). To do this, the instrument and its associated software performed a non-targeted search for deleterious substances in a dog food sample suspected of causing gastrointestinal problems in dogs. A single-stage Orbitrap/high-performance liquid chromatography method and differential expression analysis software (Sieve) was used to detect and identify, and subsequently quantify, nontargeted compounds occurring only in the suspect dog food sample. When combined with an online database (ChemSpider), a preliminary identification of one of the nontargeted compounds was determined to be potato glycoalkaloids. The diagnostic product ion ratios and quantitative data accuracy generated by the single-stage Orbitrap MS were shown to be similar to results obtained using a triple quadrupole LC-MS/MS. Additionally, the ability of the single-stage Orbitrap instrument to provide precursor and product ion accurate masses and isotope patterns was also investigated.
本研究旨在对单级 Exactive 高分辨率质谱仪(HRMS)的非靶向搜索和定量能力进行初步研究。为此,仪器及其相关软件对疑似引起犬胃肠道问题的犬粮样品中的有害物质进行了非靶向搜索。采用单级轨道阱/高效液相色谱法和差异表达分析软件(Sieve)检测和鉴定,随后对仅出现在可疑犬粮样品中的非靶向化合物进行定量分析。与在线数据库(ChemSpider)结合使用,初步确定其中一种非靶向化合物为土豆糖苷生物碱。单级轨道阱 MS 产生的诊断产物离子比和定量数据准确性与使用三重四极杆 LC-MS/MS 获得的结果相似。此外,还研究了单级轨道阱仪器提供前体和产物离子精确质量和同位素模式的能力。