Peacock John, Dirckx Joris, von Unge Magnus
Laboratory of Biomedical Physics, University of Antwerp , Antwerp , Belgium.
Acta Otolaryngol. 2015 Sep;135(9):880-5. doi: 10.3109/00016489.2015.1015605. Epub 2015 May 8.
Information on the degree of stapes fixation can be found by measuring the ratio of stapes to umbo and stapes to incus velocity.
To evaluate a method of quantifying ossicular fixation in an ear with elevated tympanic membrane.
Measurements were made on four fresh-frozen human temporal bones. After elevating the tympanic membrane, a small magnet was attached to the manubrium and an electromagnetic excitation coil was used to vibrate the ossicles. The vibration response of the umbo, the tip of the incus long process, and the posterior crus of the stapes were measured before and after partially fixing the footplate with luting cement.
The velocities at the different measurement points were unequally affected by the fixation. The difference in the velocity ratio between different points provides an indication of the degree of footplate fixation.
通过测量镫骨与鼓膜脐部以及镫骨与砧骨的速度比,可以获得镫骨固定程度的信息。
评估一种量化鼓膜抬高耳中听骨链固定的方法。
对四块新鲜冷冻的人类颞骨进行测量。抬高鼓膜后,在锤骨柄上附着一个小磁铁,并使用电磁激励线圈使听骨链振动。在用粘固剂部分固定镫骨足板前后,测量鼓膜脐部、砧骨长突尖端和镫骨后脚的振动响应。
不同测量点的速度受固定的影响程度不同。不同点之间速度比的差异可表明足板固定的程度。