We report a label free electrochemical detection of cardiac bio-marker myoglobin (Mb) on aptamer functionalized rGO/CNT nanostructured electrodes by measuring its direct electron transfer (DET). Configured as a highly responsive aptasensor, the newly developed biosensing platform exhibits synergistic effect of the nano-hybrid functional construct by combining good electrical properties and the facile chemical functionality of nanohybrid for the compatible bio-interface development. The specific anti-Mb aptamer was generated by five iterative SELEX (Systematic evolution of ligands by exponential enrichment) rounds, showing high senstivity (KD ~65 pM). The aptamer functionalized rGO/CNT nanostructured electrodes demonstrated a significant increase in signal response with a detection limit of ~0.34 ng/mL in the dynamic response range between 1 ng/mL and 4 µg/mL for Mb. The newly developed DET assay format presents a promising candidate in point-of-care diagnosis for routine screening of Mb in patient's samples.
我们报告了一种无标记的电化学检测方法,可通过测量其直接电子转移(DET),在适配体功能化 rGO/CNT 纳米结构电极上检测心脏生物标志物肌红蛋白(Mb)。作为一种高响应的适体传感器,新开发的生物传感平台通过结合纳米杂化物的良好电性能和易于化学功能化的特性,发挥了纳米杂化物功能结构的协同效应,用于兼容的生物界面开发。通过五轮迭代 SELEX(指数富集的配体系统进化),产生了特异性抗 Mb 适配体,具有高灵敏度(KD~65 pM)。在 Mb 的动态响应范围内为 1ng/mL 至 4μg/mL 之间,适配体功能化 rGO/CNT 纳米结构电极的信号响应显著增加,检测限约为 0.34ng/mL。新开发的 DET 检测方法在即时检测中具有很大的应用潜力,可用于常规筛选患者样本中的 Mb。