Early malfunction of polyvinyl alcohol membrane of septal occluder.

作者信息

Ramoğlu Mehmet G, Uçar Tayfun, Tutar Ercan

机构信息

Department of Pediatric Cardiology, Fellow in Pediatric Cardiology, Ankara University, Medical School, Ankara, Turkey.

Department of Pediatric Cardiology, Associate Professor in Pediatric Cardiology, Ankara University, Medical School, Ankara, Turkey.

出版信息

Catheter Cardiovasc Interv. 2016 Mar;87(4):E151-3. doi: 10.1002/ccd.26057. Epub 2015 Jun 23.

Abstract

Percutaneous device closure of atrial septal defect (ASD) is now considered the choice of treatment. Numerous devices with advantages/disadvantages are currently available and under development. Cardia Ultrasept II ASD occluder has a nitinol frame covered with polyvinyl alcohol (PVA) membrane. Here, a well-documented case of early malfunction of PVA membrane detected in the first week of implantation in a 4-year-old male patient, who underwent an uneventful device closure with 20 mm Cardia Ultrasept II ASD Occluder, is presented. One week after implantation left-to-right shunt through the device was detected and the explantation of device revealed PVA membrane with multiple perforations.