Suppr超能文献

142枚单一设计旋转铰链膝关节假体的中期结果:频发并发症可能不影响严重受损膝关节的挽救治疗

Intermediate-Term Results of 142 Single-Design, Rotating-Hinge Implants: Frequent Complications May Not Preclude Salvage of Severely Affected Knees.

作者信息

Farid Yasser R, Thakral Rishi, Finn Henry A

机构信息

Assistant professor, Clinical orthopedic surgery, University of Illinois at Chicago and Weiss Memorial Hospital Bone and Joint Replacement Center, Chicago, Illinois.

Fellow, University of Chicago and Weiss Memorial Hospital Bone and Joint Replacement Center, Chicago, Illinois.

出版信息

J Arthroplasty. 2015 Dec;30(12):2173-80. doi: 10.1016/j.arth.2015.06.033. Epub 2015 Jun 20.

Abstract

Rotating-hinge knee prostheses have low survivorship and high complications except in primary arthroplasties in elderly patients. We retrospectively reviewed 142 single third-generation design, rotating hinge prostheses (11 primary procedures and 131 revisions) at 57 months follow up. Implant survival was 73%. Successful two-stage reimplantation for prosthetic infection was 78.4% but new infection rate was 22%. The tibial component was durable while the femoral component was problematic. We observed only one patellar maltracking and no polyethylene wear. A third generation rotating-hinge arthroplasty reconstruction was reliable in complex problems. Outcomes in primary situations were excellent. Complications were the rule rather than the exception in revisions. With timely intervention, attention to soft tissue coverage, and realistic expectations, complications were contained and functional benefits were appreciable.

摘要

旋转铰链膝关节假体除用于老年患者的初次关节成形术外,生存率较低且并发症发生率较高。我们对142例采用第三代单髁设计的旋转铰链假体进行了回顾性研究(11例初次手术和131例翻修手术),随访57个月。植入物生存率为73%。假体感染的二期再植入成功率为78.4%,但新感染率为22%。胫骨部件耐用,而股骨部件存在问题。我们仅观察到1例髌骨轨迹不良,未发现聚乙烯磨损。第三代旋转铰链关节成形术重建在复杂问题中是可靠的。初次手术的效果极佳。翻修手术中并发症很常见而非个别现象。通过及时干预、关注软组织覆盖情况以及抱有现实的期望,并发症得到了控制,功能改善明显。