Suppr超能文献

一种评估不透射线钽髋臼杯聚乙烯内衬磨损的新方法:纳入随机对照试验患者的临床验证

A Novel Method for Assessment of Polyethylene Liner Wear in Radiopaque Tantalum Acetabular Cups: Clinical Validation in Patients Enrolled in a Randomized Controlled Trial.

作者信息

Troelsen Anders, Greene Meridith E, Ayers David C, Bragdon Charles R, Malchau Henrik

机构信息

Harris Orthopaedic Laboratory, Department of Orthopedics, Massachusetts General Hospital, Boston, Massachusetts; Department of Orthopedic Surgery, Copenhagen University Hospital Hvidovre, Hvidovre, Denmark.

Harris Orthopaedic Laboratory, Department of Orthopedics, Massachusetts General Hospital, Boston, Massachusetts.

出版信息

J Arthroplasty. 2015 Dec;30(12):2354-9. doi: 10.1016/j.arth.2015.06.050. Epub 2015 Jul 2.

Abstract

Conventional radiostereometric analysis (RSA) for wear is not possible in patients with tantalum cups. We propose a novel method for wear analysis in tantalum cups. Wear was assessed by gold standard RSA and the novel method in total hip arthroplasty patients enrolled in a randomized controlled trial receiving either titanium or tantalum cups (n=46). The novel method estimated the center of the head using a model based on identification of two proximal markers on the stem and knowledge of the stem/head configuration. The novel method was able to demonstrate a pattern of wear that was similar to the gold standard in titanium cups. The novel method offered accurate assessment and is a viable solution for assessment of wear in studies with tantalum cups.

摘要

对于使用钽杯的患者,无法进行常规的磨损放射立体测量分析(RSA)。我们提出了一种用于钽杯磨损分析的新方法。在一项随机对照试验中,对接受钛杯或钽杯的全髋关节置换术患者(n = 46),采用金标准RSA和新方法评估磨损情况。新方法基于识别柄上的两个近端标记并结合柄/头的结构模型来估计股骨头中心。新方法能够呈现出与钛杯金标准相似的磨损模式。该新方法提供了准确的评估,是钽杯研究中评估磨损的可行解决方案。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验