Troelsen Anders, Greene Meridith E, Ayers David C, Bragdon Charles R, Malchau Henrik
Harris Orthopaedic Laboratory, Department of Orthopedics, Massachusetts General Hospital, Boston, Massachusetts; Department of Orthopedic Surgery, Copenhagen University Hospital Hvidovre, Hvidovre, Denmark.
Harris Orthopaedic Laboratory, Department of Orthopedics, Massachusetts General Hospital, Boston, Massachusetts.
J Arthroplasty. 2015 Dec;30(12):2354-9. doi: 10.1016/j.arth.2015.06.050. Epub 2015 Jul 2.
Conventional radiostereometric analysis (RSA) for wear is not possible in patients with tantalum cups. We propose a novel method for wear analysis in tantalum cups. Wear was assessed by gold standard RSA and the novel method in total hip arthroplasty patients enrolled in a randomized controlled trial receiving either titanium or tantalum cups (n=46). The novel method estimated the center of the head using a model based on identification of two proximal markers on the stem and knowledge of the stem/head configuration. The novel method was able to demonstrate a pattern of wear that was similar to the gold standard in titanium cups. The novel method offered accurate assessment and is a viable solution for assessment of wear in studies with tantalum cups.