Development of two-band infrared radiometer for irradiance calibration of target simulators.

作者信息

Yang Sen, Li Chengwei

机构信息

School of Electrical Engineering and Automation, Harbin Institute of Technology, 150001 Harbin, People's Republic of China.

出版信息

Rev Sci Instrum. 2015 Jul;86(7):073105. doi: 10.1063/1.4926347.

Abstract

A detector-based spectral radiometer has been developed for the calibration of target simulator. Unlike the conventional spectral irradiance calibration method based on radiance and irradiance, the new radiometer is calibrated using image-space temperature based method. The image-space temperature based method improves the reproducibility in the calibration of radiometer and reduces the uncertainties existing in the conventional calibration methods. The calibrated radiometer is then used to establish the irradiance transfer standard for the target simulator. With the designed radiometer in this paper, a highly accurate irradiance calibration for target simulators of wavelength from 2.05 to 2.55 μm and from 3.7 to 4.8 μm can be performed with an expanded uncertainty (k = 2) of calibration of 2.18%. Last but not least, the infrared radiation of the target simulator was measured by the infrared radiometer, the effectiveness and capability of which are verified through measurement of temperature and irradiance and a comparison with the thermal imaging camera.

摘要

已开发出一种基于探测器的光谱辐射计用于目标模拟器的校准。与基于辐射亮度和辐照度的传统光谱辐照度校准方法不同,这种新型辐射计采用基于图像空间温度的方法进行校准。基于图像空间温度的方法提高了辐射计校准的可重复性,并减少了传统校准方法中存在的不确定性。然后,使用经过校准的辐射计为目标模拟器建立辐照度传递标准。利用本文设计的辐射计,可对波长在2.05至2.55μm以及3.7至4.8μm范围内的目标模拟器进行高精度辐照度校准,校准的扩展不确定度(k = 2)为2.18%。最后,用红外辐射计测量了目标模拟器的红外辐射,通过温度和辐照度测量以及与热成像相机的比较,验证了其有效性和性能。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验