Suppr超能文献

一种提高离子射线照相空间分辨率的先进图像处理方法。

An advanced image processing method to improve the spatial resolution of ion radiographies.

作者信息

Krah N, Testa M, Brons S, Jäkel O, Parodi K, Voss B, Rinaldi I

机构信息

Heidelberg Collaboratory for Image Processing, Speyerer Str. 6, 69115 Heidelberg, Germany.

出版信息

Phys Med Biol. 2015 Nov 7;60(21):8525-47. doi: 10.1088/0031-9155/60/21/8525. Epub 2015 Oct 20.

Abstract

We present an optimization method to improve the spatial resolution and the water equivalent thickness (WET) accuracy of ion radiographies. The method is designed for imaging systems measuring for each actively scanned beam spot the lateral position of the pencil beam and at the same time the Bragg curve (behind the target) in discrete steps without relying on tracker detectors to determine the ion trajectory before and after the irradiated volume. Specifically, the method was used for an imaging set-up consisting of a stack of 61 parallel-plate ionization chambers (PPIC) interleaved with absorber plates of polymethyl methacrylate (PMMA) working as a range telescope. The method uses not only the Bragg peak position, but approximates the entire measured Bragg curve as a superposition of differently shifted Bragg curves. Their relative weights allow to reconstruct the distribution of thickness around each scan spot of a heterogeneous phantom. The approach also allows merging the ion radiography with the geometric information of a co-registered x-ray radiography in order to increase its spatial resolution. The method was tested using Monte Carlo simulated and experimental proton radiographies of a PMMA step phantom and an anthropomorphic head phantom. For the step phantom, the effective spatial resolution was found to be 6 and 4 times higher than the nominal resolution for the simulated and experimental radiographies, respectively. For the head phantom, a gamma index was calculated to quantify the conformity of the simulated proton radiographies with a digitally reconstructed radiography (DRR) obtained from an x-ray CT and properly converted into WET. For a distance-to-agreement (DTA) of 2.5 mm and a relative WET difference (RWET) of 2.5%, the passing ratio was 100%/85% for the optimized/non-optimized case, respectively. When the optimized proton radiography was merged with the co-registered DRR, the passing ratio was 100% at DTA  =  1.3 mm and RWET  =  1.3%. A special interpolation method allows to strongly reduce the dose by using a coarser grid of the measured beam spot position with a 5 times larger grid distance. We show that despite a dose reduction of 25 times (leading to a dose of 0.016 mGy for the current imaging set-up), the image quality of the optimized radiographies remains fairly unaffected for both the simulated and experimental case.

摘要

我们提出了一种优化方法,以提高离子射线成像的空间分辨率和水等效厚度(WET)精度。该方法适用于成像系统,该系统在不依赖跟踪探测器来确定离子在照射体积前后轨迹的情况下,以离散步骤测量每个主动扫描束斑的铅笔束横向位置以及同时测量(靶标后方的)布拉格曲线。具体而言,该方法用于一种成像装置,该装置由61个平行板电离室(PPIC)堆叠而成,中间夹有聚甲基丙烯酸甲酯(PMMA)吸收板,用作射程望远镜。该方法不仅使用布拉格峰位置,还将整个测量的布拉格曲线近似为不同偏移的布拉格曲线的叠加。它们的相对权重允许重建异质体模每个扫描点周围的厚度分布。该方法还允许将离子射线成像与配准的X射线成像的几何信息合并,以提高其空间分辨率。该方法使用蒙特卡罗模拟以及PMMA阶梯体模和人体头部体模的实验质子射线成像进行了测试。对于阶梯体模,发现模拟和实验射线成像的有效空间分辨率分别比标称分辨率高6倍和4倍。对于头部体模,计算了伽马指数,以量化模拟质子射线成像与从X射线CT获得并适当转换为WET的数字重建射线成像(DRR)的符合程度。对于2.5毫米的一致性距离(DTA)和2.5%的相对WET差异(RWET),优化/未优化情况下的通过率分别为100%/85%。当优化的质子射线成像与配准的DRR合并时,在DTA = 1.3毫米和RWET = 1.3%时通过率为100%。一种特殊的插值方法允许通过使用测量束斑位置的更粗网格(网格距离大5倍)来大幅降低剂量。我们表明,尽管剂量降低了25倍(导致当前成像装置的剂量为0.016毫戈瑞),但优化射线成像的图像质量在模拟和实验情况下仍基本不受影响。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验