Suppr超能文献

Comparison of fluid leakage across endotracheal tube cuffs using a three-dimensional printed model of the human trachea.

作者信息

Kimijima Tomohiko, Edanaga Mitsutaka, Yamakage Michiaki

机构信息

Department of Anesthesiology, Sapporo Medical University School of Medicine, S 1, W 16, Chuo-ku, Sapporo, Hokkaido, 060-8543, Japan.

出版信息

J Anesth. 2016 Jun;30(3):510-3. doi: 10.1007/s00540-016-2138-9. Epub 2016 Jan 27.

Abstract

The objective of this study was to compare fluid leakage across endotracheal tube cuffs using a three-dimensional (3D)-printed human tracheal model that anatomically simulates the human trachea. We made two models based on computed tomography data of the neck and chest. Using a Mallinckrodt Hi-Lo™ (HL), ThinCuff(®) (TC), and Mallinckrodt TaperGuard™ (TG), we sequentially measured the amount of fluid leakage across each endotracheal tube cuff after applying saline or viscous liquid above the cuff. The TG allowed significantly less leakage than the HL and TC with both saline and the viscous liquid. Our study, using a 3D-printed tracheal model, indicated that a conical-shaped endotracheal tube cuff significantly reduces fluid leakage across the cuff compared with conventional cylindrical-shaped cuffs made of polyurethane or polyvinylchloride, contrary to the results of a previous study using a solid cylindrical structure.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验