This article reviews and explains the basic physical principles of metal-induced MRI artifacts, describes simple ways to reduce them, and presents specific reduction solutions. Artifacts include signal loss, pile-up artifacts, geometric distortion, and failure of fat suppression. Their nature and origins are reviewed and explained though schematic representations that ease the understanding. Then, optimization of simple acquisition parameters is detailed. Lastly, dedicated sequences and options specifically developed to reduce metal artifacts (VAT, SEMAC, and MAVRIC) are explained.
本文回顾并解释了金属诱导的磁共振成像伪影的基本物理原理,描述了减少这些伪影的简单方法,并给出了具体的减少方案。伪影包括信号丢失、堆积伪影、几何畸变和脂肪抑制失败。通过便于理解的示意图对它们的性质和起源进行了回顾和解释。然后详细介绍了简单采集参数的优化。最后,解释了专门为减少金属伪影而开发的特定序列和选项(VAT、SEMAC和MAVRIC)。