Targeted Drug Delivery Research Center, School of Pharmacy, Mashhad University of Medical Sciences, Mashhad, Iran.
Nanotechnology Research Center, School of Pharmacy, Mashhad University of Medical Sciences, Mashhad, Iran; Research Institute of Sciences and New Technology, Mashhad, Iran.
Biosens Bioelectron. 2016 Jun 15;80:532-537. doi: 10.1016/j.bios.2016.02.029. Epub 2016 Feb 11.
Monitoring of myoglobin (Mb) in human blood serum is highly in demand for early diagnosis of acute myocardial infarction (AMI). Here, a novel electrochemical aptasensor was developed for ultrasensitive and selective detection of Mb, based on Y-shape structure of dual-aptamer (DApt)-complementary strand of aptamer (CS) conjugate, gold electrode and exonuclease I (Exo I). The designed aptasensor obtains features of gold, such as high electrochemical conductivity and large surface area, property of Y-shape structure of DApt-CS conjugate to function as a gate and obstacle for the access of redox probe to the surface of electrode, as well as high specificity and sensitivity of aptamer toward its target and Exo I as an enzyme which specifically degrades the 3'-end of single-stranded DNA (ssDNA). In the absence of Mb, the Y-shape structure remains intact. So, a weak electrochemical signal is observed. Upon addition of target, the DApt leave the CS and bind to Mb, leading to disassembly of Y-shape structure and following the addition of Exo I, a strong electrochemical signal could be recorded. The fabricated aptasensor showed high selectivity toward Mb with a limit of detection (LOD) as low as 27 pM. Besides, the developed aptasensor was effectively applied to detect Mb in human serum.
监测人血清中的肌红蛋白(Mb)对于急性心肌梗死(AMI)的早期诊断具有很高的需求。在这里,我们基于双适体(DApt)-适体互补链(CS)缀合物的 Y 型结构、金电极和核酸外切酶 I(Exo I),开发了一种用于 Mb 超灵敏和选择性检测的新型电化学适体传感器。所设计的适体传感器具有金的高导电性和大表面积等特性、Y 型结构的 DApt-CS 缀合物作为门和障碍物的功能,可阻止氧化还原探针进入电极表面、适体对其靶标和 Exo I 的高特异性和灵敏度,Exo I 是一种特异性降解单链 DNA(ssDNA)3'-端的酶。在没有 Mb 的情况下,Y 型结构保持完整。因此,观察到微弱的电化学信号。加入靶标后,DApt 离开 CS 并与 Mb 结合,导致 Y 型结构解体,随后加入 Exo I,可记录到强烈的电化学信号。所制备的适体传感器对 Mb 具有高选择性,检测限(LOD)低至 27 pM。此外,该适体传感器还可有效用于检测人血清中的 Mb。