Coupling of an active middle-ear implant to the long process of the incus using an elastic clip attachment.

作者信息

Schraven Sebastian P, Mlynski Robert, Dalhoff Ernst, Heyd Andreas, Wildenstein Daniela, Rak Kristen, Radeloff Andreas, Hagen Rudolf, Gummer Anthony W

机构信息

Department of Oto-Rhino-Laryngology, Plastic, Aesthetic and Reconstructive Head and Neck Surgery, Comprehensive Hearing Center, University of Würzburg, Josef-Schneider-Straße 11, D-97080 Würzburg, Germany.

Department of Otorhinolaryngology, Head and Neck Surgery "Otto Körner", University Medical Center Rostock, Doberaner Straße 137-139, D-18057 Rostock, Germany.

出版信息

Hear Res. 2016 Oct;340:179-184. doi: 10.1016/j.heares.2016.03.012. Epub 2016 Apr 13.

Abstract

The active middle-ear implant Vibrant Soundbridge (VSB) is used to treat mild-to-severe sensorineural hearing losses. The standard surgical approach for incus vibroplasty is a mastoidectomy and a posterior tympanotomy, crimping the Floating Mass Transducer (FMT) to the long process of the incus (LPI) (standard crimped application). However, tight crimping increases the risk of necrosis of the LPI, resulting in reduction of energy transfer and loss of amplification. The aim of this study was to develop a new coupling device for the LPI, that does not require crimping, and to test its vibrational transfer properties in temporal-bone preparations. An extended antrotomy and a posterior tympanotomy were performed in ten fresh human temporal bones. As a control for normal middle-ear function, the tympanic membrane was stimulated acoustically and the vibration of the stapes footplate was measured by laser Doppler vibrometry (LDV). FMT-induced vibration responses of the stapes were then measured for the standard crimped application at the LPI and for the newly designed elastic long process coupler (LP coupler). For the LP coupler, velocity-amplitude responses in temporal-bone preparations showed increased mean amplitudes at around 1 kHz (∼10 dB) and a reduction between 1.8 and 6 kHz (13 dB on average for 2 ≤ f ≤ 5 kHz). In conclusion, attachment of the FMT to the LPI with the LP coupler leads to generally good mechanical and functional coupling in temporal-bone preparations with a notable disadvantage between 1.8 and 6 kHz. Due to its elastic clip attachment it is expected that the LP coupler will reduce the risk of necrosis of the incus long process, which has to been shown in further studies. Clinical results of the LP coupler are pending.

摘要

有源中耳植入物“活力声桥”(VSB)用于治疗轻至重度感音神经性听力损失。砧骨振动成形术的标准手术方法是乳突切除术和后鼓室切开术,将浮动质量传感器(FMT)压接至砧骨长突(LPI)(标准压接应用)。然而,紧密压接会增加LPI坏死的风险,导致能量传递减少和放大功能丧失。本研究的目的是开发一种用于LPI的新型耦合装置,该装置无需压接,并在颞骨标本中测试其振动传递特性。对10块新鲜人颞骨进行了扩大的鼓窦切开术和后鼓室切开术。作为正常中耳功能的对照,通过声学刺激鼓膜,并使用激光多普勒振动测量法(LDV)测量镫骨足板的振动。然后分别测量LPI处标准压接应用以及新设计的弹性长突耦合器(LP耦合器)时FMT引起的镫骨振动响应。对于LP耦合器,颞骨标本中的速度-振幅响应显示,在1kHz左右平均振幅增加(约10dB),在1.8至6kHz之间降低(2≤f≤5kHz时平均降低13dB)。总之,使用LP耦合器将FMT附着于LPI,在颞骨标本中通常能实现良好的机械和功能耦合,但在1.8至6kHz之间存在明显劣势。由于其弹性夹固定方式,预计LP耦合器将降低砧骨长突坏死的风险,这有待进一步研究证实。LP耦合器的临床结果尚待确定。