J Cataract Refract Surg. 2016 Apr;42(4):577-81. doi: 10.1016/j.jcrs.2016.01.038.
To assess the repeatability of a new swept-source optical coherence tomography (SS-OCT) biometer (IOLMaster 700) and evaluate the agreement with an optical low-coherence reflectometry (OLCR) biometer (Lenstar 900) in cataractous eyes.
Tertiary eyecare facility.
Prospective evaluation of a diagnostic test.
Eyes of patients had 3 consecutive scans acquired by the same operator. The repeatability of measurements was assessed using the within-subject standard deviation and coefficient of variation (CoV). Agreement between the mean measurements of each machine was evaluated.
The study comprised 100 eyes of 100 patients. The median lens thickness and axial length (AL) measurements were significantly greater (P < .001) with SS-OCT (4.29 mm and 23.5 mm, respectively, versus 4.25 mm and 23.4 mm, respectively); anterior chamber depth (ACD) was greater with OLCR (2.72 mm versus 2.70 mm, respectively) (P < .001). The within-subject standard deviation and CoV of the AL (0.01 and 0.05, respectively) and the ACD (0.04 and 1.22, respectively) measurements showed lower variability with SS-OCT (0.05 and 0.21, respectively, for AL; 0.06 and 1.99, respectively, for ACD) than with OLCR. Good agreement was seen for anterior segment measurements between the 2 devices. The 95% limits of agreement of most parameters fell within the test-retest variability of the parameters. The SS-OCT biometer also showed better penetration in dense posterior subcapsular cataracts, measuring AL successfully in 96% of cases.
The new SS-OCT biometer showed good repeatability and agreement with the OLCR biometer. The SS-OCT biometer measured the AL with fewer dropouts compared with the OLCR even in dense cataracts.
None of the authors has a financial or proprietary interest in any material or method mentioned.
评估新型扫频源光相干断层扫描仪(SS-OCT)生物测量仪(IOLMaster 700)的重复性,并评估其在白内障眼中与光学低相干反射计(OLCR)生物测量仪(Lenstar 900)的一致性。
三级眼科医疗机构。
诊断测试的前瞻性评估。
由同一位操作者对患者的眼睛进行 3 次连续扫描。使用个体内标准差和变异系数(CoV)评估测量的重复性。评估每台机器的平均测量值之间的一致性。
该研究包括 100 例 100 只眼。SS-OCT 的晶状体厚度和眼轴(AL)中位数测量值显著更大(P<0.001),分别为 4.29mm 和 23.5mm,而分别为 4.25mm 和 23.4mm;前房深度(ACD)OLCR 更大(2.72mm 与 2.70mm,分别)(P<0.001)。AL(0.01 和 0.05,分别)和 ACD(0.04 和 1.22,分别)测量值的个体内标准差和 CoV 显示,SS-OCT 的变异性更低(AL 分别为 0.05 和 0.21,ACD 分别为 0.06 和 1.99),与 OLCR 相比。两种设备的前节测量值之间具有良好的一致性。大多数参数的 95%界限均在参数的测试-再测试变异性内。SS-OCT 生物测量仪在密集的后囊下白内障中也具有更好的穿透力,成功测量了 96%的 AL。
新型 SS-OCT 生物测量仪具有良好的重复性和与 OLCR 生物测量仪的一致性。与 OLCR 相比,SS-OCT 生物测量仪即使在密集白内障中也能更频繁地测量 AL,并且没有测量失败。
无作者在文中提到的任何材料或方法中拥有财务或专有权益。