Suppr超能文献

一种新型光学生物测量仪与眼前节光学相干断层扫描仪在测量中央角膜厚度和前房深度方面的比较。

Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth.

作者信息

Huang Jinhai, Lu Weicong, Savini Giacomo, Chen Hao, Wang Chengfang, Yu Xinxin, Bao Fangjun, Wang Qinmei

机构信息

School of Ophthalmology and Optometry, Wenzhou Medical University, Wenzhou, Zhejiang, China; Key Laboratory of Vision Science, Ministry of Health, Wenzhou, Zhejiang, China.

G.B. Bietti Foundation IRCCS, Rome, Italy.

出版信息

J Ophthalmol. 2016;2016:6347236. doi: 10.1155/2016/6347236. Epub 2016 Jun 14.

Abstract

Purpose. To compare between a new optical biometer (AL-Scan, Nidek Co., Aichi, Japan) and an anterior segment optical coherence tomographer (Visante AS-OCT, Carl Zeiss Meditec, Dublin, USA) for measuring central corneal thickness (CCT), anterior chamber depth (ACD), and aqueous depth (AD). Methods. Sixty-three eyes of 63 normal subjects were examined with AL-Scan and Visante AS-OCT in this prospective study. One eye per subject was measured three times with both devices to record their CCT, ACD, and AD. All procedures were performed by the same operator. Agreement between the two devices was assessed using paired t-tests, Bland-Altman plots, and 95% limits of agreement (LoA). Results. The mean CCT, ACD, and AD measured by AL-Scan were 538.59 ± 27.37 μm, 3.70 ± 0.30 mm, and 3.16 ± 0.30 mm, respectively. The mean values obtained by the Visante OCT were 536.14 ± 26.61 μm for CCT, 3.71 ± 0.29 mm for ACD, and 3.17 ± 0.29 mm for AD. The mean CCT by the AL-Scan was higher than that obtained by the Visante AS-OCT (difference = 2.45 ± 6.07 μm, P < 0.05). The differences in ACD and AD measurements were not statistically significant. The 95% LoA of CCT, ACD, and AD were between -9.44 and 14.35 μm, -0.15 and 0.12 mm, and -0.15 and 0.12 mm, respectively. Conclusions. Since these two devices were comparable for measuring CCT, ACD, and AD, their results can be interchangeably used in the clinic.

摘要

目的。比较一种新型光学生物测量仪(AL-Scan,日本爱知县尼德克公司)和眼前节光学相干断层扫描仪(Visante AS-OCT,美国都柏林卡尔蔡司医疗技术公司)在测量中央角膜厚度(CCT)、前房深度(ACD)和房水深度(AD)方面的差异。方法。在这项前瞻性研究中,使用AL-Scan和Visante AS-OCT对63名正常受试者的63只眼睛进行了检查。每位受试者的一只眼睛用两种设备各测量三次,以记录其CCT、ACD和AD。所有操作均由同一名操作人员进行。使用配对t检验、Bland-Altman图和95%一致性界限(LoA)评估两种设备之间的一致性。结果。AL-Scan测量的平均CCT、ACD和AD分别为538.59±27.37μm、3.70±0.30mm和3.16±0.30mm。Visante OCT获得的平均值为CCT 536.14±26.61μm、ACD 3.71±0.29mm和AD 3.17±0.29mm。AL-Scan测量的平均CCT高于Visante AS-OCT获得的结果(差值=2.45±6.07μm,P<0.05)。ACD和AD测量的差异无统计学意义。CCT、ACD和AD的95%LoA分别在-9.44至14.35μm、-0.15至0.12mm和-0.15至0.12mm之间。结论。由于这两种设备在测量CCT、ACD和AD方面具有可比性,其结果在临床上可互换使用。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/800f/4923600/99afb232258f/JOPH2016-6347236.001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验