[Comparison of the Results of the Flicker Perimetry (Prototype versus Commercial Perimeter)].

作者信息

Zimmermann N, Hess A, Erb C

机构信息

Augenheilkunde, Augenklinik am Wittenbergplatz, Berlin.

出版信息

Klin Monbl Augenheilkd. 2017 Sep;234(9):1169-1173. doi: 10.1055/s-0042-109706. Epub 2016 Aug 10.

Abstract

Analysis of the comparison of flicker perimetry with the prototype Pulsar and the new Octopus 600. Both eyes of 20 patients with primary open-angle glaucoma were studied with 30° visual field flimmer perimetry (dynamic strategy) at the perimeters Pulsar and Octopus 600 (Haag-Streit). The evaluation was based on the mean defect (MD = mean deviation) and the defect depth (square root of lost variance = sLV). In the t test for paired samples, the Pulsar perimetry showed significantly higher values than the Octopus 600 perimetry: right eye MD 4.8 ± 3.6 src vs. 1.7 ± 2.9 src, p = 0.005; left eye MD 3.9 ± 3.6 src v. 1.4 ± 2.8 src, p = 0.018; both eyes MD 4.35 ± 3.62 src vs. 1.55 ± 2.80 src, p = 0.002. The sLV values with the Pulsar perimetry were significantly higher than the values with the Octopus 600 perimetry: right eye sLV 3.6 ± 1.6 vs. 2.3 ± 1.3 src, p = 0.006; left eye sLV 3.2 ± 0.8 vs. 2.0 ± 0.8 src, p < 0.0001; both eyes sLV 3.37 ± 1.28 src vs. 2.12 ± 1.05 src, p < 0.0001. The significances even persisted after the Bonferroni-Holm correction. The investigation results of flicker perimetry are not comparable, because there is a significant difference in MD and sLV. This mainly due to different light intensities, background brightness and varying colour scalings of the perimetric device.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验