Pakvis Dean F M, Heesterbeek Petra J C, Severens Marianne, Spruit Maarten
a Department of Orthopaedic surgery , Sint Maartenskliniek.
b Sint Maartenskliniek Research , Nijmegen , the Netherlands.
Acta Orthop. 2016 Dec;87(6):583-588. doi: 10.1080/17453674.2016.1237439. Epub 2016 Sep 23.
Background and purpose - The acetabular component has remained the weakest link in hip arthroplasty for achievement of long-term survival. One of the possible explanatory factors for acetabular failure has been acetabular stress shielding. For this, we investigated the effects of a cementless elastic socket on acetabular bone mineral density (BMD). Patients and methods - During 2008-2009, we performed a single-center prospective cohort trial on 25 patients (mean age 64 (SD 4), 18 females) in whom we implanted a cementless elastic press-fit socket. Using quantitative BMD measurements on CT, we determined the change in BMD surrounding the acetabular component over a 2-year follow-up period. Results - We found a statistically significant decrease in cancellous BMD (-14% to -35%) and a stable level of cortical BMD (5% to -5%) surrounding the elastic press-fit cup during the follow-up period. The main decrease was seen during the first 6 months after implantation. During the second year, cancellous BMD showed a further decrease in the medial and lower acetabular regions. Interpretation - We found no evidence that an elastic press-fit socket would prevent acetabular stress shielding during a 2-year follow-up.
背景与目的——在髋关节置换术中,髋臼假体一直是影响长期生存率的最薄弱环节。髋臼应力遮挡是导致髋臼假体失败的可能原因之一。为此,我们研究了非骨水泥弹性髋臼杯对髋臼骨密度(BMD)的影响。
患者与方法——在2008年至2009年期间,我们对25例患者(平均年龄64岁(标准差4),18例女性)进行了单中心前瞻性队列研究,为这些患者植入了非骨水泥弹性压配式髋臼杯。通过CT定量测量骨密度,我们确定了在2年随访期内髋臼假体周围骨密度的变化。
结果——我们发现,在随访期间,弹性压配式髋臼杯周围的松质骨骨密度有统计学意义的下降(-14%至-35%),皮质骨骨密度保持稳定(5%至-5%)。主要下降发生在植入后的前6个月。在第二年,髋臼中下部区域的松质骨骨密度进一步下降。
解读——我们没有发现证据表明弹性压配式髋臼杯在2年随访期内可预防髋臼应力遮挡。