Aikenhead J, Triano J, Baker J
Department of Clinical Radiology, National College of Chiropractic, Lombard, IL 60148.
J Manipulative Physiol Ther. 1989 Aug;12(4):259-64.
Radiation dosages to sensitive organs in full spine radiography have in recent years been a concern of physicians as well as the general public. The spine is the prime target for exposure in scoliosis radiography, though the exposure usually necessitates irradiation of several radio-sensitive organs. In recent studies, various protection techniques have been used including various lead and aluminum filtration systems, altered patient positioning and varied tube-film distances. The purpose of this study was to evaluate the efficiency for radiation dosage reduction of three filtration systems used frequently in the chiropractic profession. The systems tested were the Nolan Multiple X-ray Filters, the Clear-Pb system and the Sportelli Wedge system. These systems were tested in seven configurations varying breast shielding, distance and patient positioning. All systems tested demonstrated significant radiation reductions to organs, especially breast tissue. The Clear-Pb system appeared to be the most effective for all organs except the breast, and the Sportelli Wedge system demonstrated the greatest reduction to breast tissue.
近年来,全脊柱X线摄影中对敏感器官的辐射剂量一直是医生和公众关注的问题。脊柱是脊柱侧弯X线摄影的主要照射部位,尽管这种照射通常需要对多个放射敏感器官进行辐照。在最近的研究中,已经使用了各种防护技术,包括各种铅和铝过滤系统、改变患者体位以及改变管-胶片距离。本研究的目的是评估脊椎按摩治疗行业中常用的三种过滤系统降低辐射剂量的效率。测试的系统是诺兰多重X射线过滤器、Clear-Pb系统和斯波泰利楔形系统。这些系统在七种配置下进行了测试,包括不同的乳房屏蔽、距离和患者体位。所有测试系统都显示出对器官,尤其是乳腺组织的辐射显著减少。Clear-Pb系统似乎对除乳房外的所有器官最有效,而斯波泰利楔形系统对乳腺组织的减少效果最大。