Lewin Andrew C, Miller Paul E
School of Veterinary Medicine, University of Wisconsin-Madison, 2015 Linden Drive, Madison, WI, 53706, USA.
Vet Ophthalmol. 2017 Nov;20(6):571-573. doi: 10.1111/vop.12445. Epub 2016 Nov 24.
The pig has an increasingly important role in ocular drug delivery models, but the most accurate tonometer in this species is unknown. The purpose of this study was to evaluate the accuracy of TonoVet and Tono-Pen Vet tonometers in the ex vivo porcine eye.
Four freshly enucleated normal porcine eyes were cannulated with two 25-gauge needles; one connected via tubing to a mercury manometer calibrated continuous physiologic recorder and the other connected to a reservoir of lactated Ringer's solution on an adjustable stand. Triplicate IOP readings were taken with the TonoVet and then the Tono-Pen Vet at 5, 10, 15, 20, 25, 30, 35, 40, 50, 60, 70, 80 mmHg.
Linear regression showed strong linear trends for both the TonoVet (r = 0.969) and Tono-Pen Vet (r = 0.983). The TonoVet slightly underestimated IOP at lower pressures and slightly overestimated IOP at higher pressures (y = 1.092x - 4.0, where y = tonometer reading, x = manometer reading, and 4.0 = intercept). The Tono-Pen Vet consistently underestimated IOP (y = 0.773x - 2.1). These differences were statistically significant (P = <0.001, one-way repeated-measures ANOVA).
As in other species, both the TonoVet and Tono-Pen Vet tonometers do not measure true IOP in the porcine eye; however, the TonoVet more closely approximated true IOP in the normal porcine eye than the Tono-Pen Vet and may be the tonometer of choice for this species.
猪在眼部给药模型中的作用日益重要,但该物种中最精确的眼压计尚不清楚。本研究的目的是评估TonoVet眼压计和Tono-Pen Vet眼压计在离体猪眼中的准确性。
用两根25号针头插入四只新鲜摘除的正常猪眼;一根通过管道连接到校准的连续生理记录仪的水银压力计,另一根连接到可调支架上的乳酸林格氏液储液器。使用TonoVet眼压计,然后使用Tono-Pen Vet眼压计在5、10、15、20、25、30、35、40、50、60、70、80 mmHg下进行三次眼压读数。
线性回归显示TonoVet眼压计(r = 0.969)和Tono-Pen Vet眼压计(r = 0.983)均有很强的线性趋势。TonoVet眼压计在较低压力下略低估眼压,在较高压力下略高估眼压(y = 1.092x - 4.0,其中y =眼压计读数,x =压力计读数,4.0 =截距)。Tono-Pen Vet眼压计始终低估眼压(y = 0.773x - 2.1)。这些差异具有统计学意义(P = <0.001,单向重复测量方差分析)。
与其他物种一样,TonoVet眼压计和Tono-Pen Vet眼压计在猪眼中均不能测量真实眼压;然而,在正常猪眼中,TonoVet眼压计比Tono-Pen Vet眼压计更接近真实眼压,可能是该物种的首选眼压计。