Suppr超能文献

A novel technique for real-time estimation of edge pedestal density gradients via reflectometer time delay data.

作者信息

Zeng L, Doyle E J, Rhodes T L, Wang G, Sung C, Peebles W A, Bobrek M

机构信息

Physics and Astronomy Department, University of California, Los Angeles, California 90095, USA.

Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831-6006, USA.

出版信息

Rev Sci Instrum. 2016 Nov;87(11):11E719. doi: 10.1063/1.4961289.

Abstract

A new model-based technique for fast estimation of the pedestal electron density gradient has been developed. The technique uses ordinary mode polarization profile reflectometer time delay data and does not require direct profile inversion. Because of its simple data processing, the technique can be readily implemented via a Field-Programmable Gate Array, so as to provide a real-time density gradient estimate, suitable for use in plasma control systems such as envisioned for ITER, and possibly for DIII-D and Experimental Advanced Superconducting Tokamak. The method is based on a simple edge plasma model with a linear pedestal density gradient and low scrape-off-layer density. By measuring reflectometer time delays for three adjacent frequencies, the pedestal density gradient can be estimated analytically via the new approach. Using existing DIII-D profile reflectometer data, the estimated density gradients obtained from the new technique are found to be in good agreement with the actual density gradients for a number of dynamic DIII-D plasma conditions.