通过荧光透视法测量并由Respitrace得出的膈肌移位。

Diaphragmatic displacement measured by fluoroscopy and derived by Respitrace.

作者信息

Verschakelen J A, Deschepper K, Jiang T X, Demedts M

机构信息

Department of Radiology, University Hospitals, Leuven, Belgium.

出版信息

J Appl Physiol (1985). 1989 Aug;67(2):694-8. doi: 10.1152/jappl.1989.67.2.694.

Abstract

In eight healthy volunteers we simultaneously measured the axial diaphragmatic motion by fluoroscopy and the cross-sectional area changes of the rib cage (RC) and abdomen (ABD) by Respitrace (RIP) during semistatic vital capacities (VC). We found that, if the fluoroscopic axial displacement of the posterior part of the diaphragm between residual volume (RV) and total lung capacity (TLC) is considered equal to 100%, the movement of the middle part is 90%, whereas that of the anterior part is only approximately 60%; the ratio of the axial displacements to mouth volume, furthermore, decreases at high lung volumes, especially for the anterior part. The RIP signal is nearly linearly related to mouth volume, but the contribution of the RC (delta RC) progressively increases (and is approximately 80% RIP at TLC), whereas the volume contribution of the ABD (delta ABD) levels off (to 20% RIP at TLC). The diaphragmatic volume displacement calculated from the theoretical analysis described by Mead and Loring also levels off at high volumes similarly as the ABD but is approximately 50% RIP at TLC. Finally, the axial movements of the three parts of the diaphragm are linearly related to the RC and ABD cross-sectional-area changes (r 0.91-0.97) and are even significantly better correlated with the "calculated" diaphragmatic volume displacement.

摘要

在八名健康志愿者中,我们在半静态肺活量(VC)期间,通过荧光透视法同步测量了膈肌的轴向运动,并通过呼吸运动描记器(Respitrace,RIP)测量了胸廓(RC)和腹部(ABD)的横截面积变化。我们发现,如果将膈肌后部在残气量(RV)和肺总量(TLC)之间的荧光透视轴向位移视为100%,那么中部的位移为90%,而前部的位移仅约为60%;此外,轴向位移与口腔容积的比值在高肺容积时会降低,尤其是前部。RIP信号与口腔容积几乎呈线性相关,但胸廓的贡献(△RC)逐渐增加(在TLC时约为RIP的80%),而腹部的容积贡献(△ABD)趋于平稳(在TLC时为RIP的20%)。根据Mead和Loring描述的理论分析计算出的膈肌容积位移在高容积时也与腹部类似地趋于平稳,但在TLC时约为RIP的50%。最后,膈肌三个部分的轴向运动与胸廓和腹部横截面积变化呈线性相关(r = 0.91 - 0.97),并且与“计算出的”膈肌容积位移的相关性甚至更好。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验