Walker Daniel T, Phillips Karen P
St Jude Medical,17 Orion Rd, Lane Cove, NSW, Australia 2066.
Heart Care Partners,Greenslopes Private Hospital,Newdegate St, Greenslopes,QLD, Australia 4120.
J Atr Fibrillation. 2014 Jun 30;7(1):1028. doi: 10.4022/jafib.1028. eCollection 2014 Jun-Jul.
The Contact (St Jude Medical) System uses a novel impedance- based measure of Electrical Coupling Index (ECI) to assess the quality of catheter tip to endocardium contact. We sought to establish average ECI measurements and behaviour during pulmonary vein (PV) isolation procedures. Forty-five patients undergoing PV isolation for atrial fibrillation (AF) were studied.'Non-contact' and upper range 'in-contact' catheter positioning was performed for system calibration. ECI measurements were recorded pre-ablation at 14 standardized locations around the PV antra. The mean ECI non-contact value was 77 ± 11 (range 63-107); the mean upper range in-contact value was 111 ± 16 (range 81-145). Mean ECI values pre-ablation around the PV antra ranged from 85 ± 18 to 107 ± 19. A trend towards higher mean ECI values was noted with increasing body mass index (BMI). Pre-ablation mean ECI values were 92 ± 10 (BMI 20-25), 95 ± 12 (BMI 26-30) and 104 ± 11 (BMI >30) (p< 0.01 for 20-25 vs. >30). A positive correlation was noted for mean pre-ablation ECI values and BMI (r=0.50). An expected range of ECI values during PV isolation has been documented in this study. Observed ECI values correlate with patient BMI. The potential limitations of the current generation Contact System and scope for future clinical applications are discussed.
Contact(圣犹达医疗)系统采用一种基于新型阻抗的电耦合指数(ECI)测量方法来评估导管尖端与心内膜接触的质量。我们试图确定肺静脉(PV)隔离手术期间的平均ECI测量值及变化情况。对45例因心房颤动(AF)接受PV隔离的患者进行了研究。为进行系统校准,采用了“非接触”和上限“接触”导管定位。在PV窦周围14个标准化位置记录消融前的ECI测量值。平均非接触ECI值为77±11(范围63 - 107);平均上限接触值为111±16(范围81 - 145)。PV窦周围消融前的平均ECI值范围为85±18至107±19。随着体重指数(BMI)增加,平均ECI值有升高趋势。消融前平均ECI值分别为92±10(BMI 20 - 25)、95±12(BMI 26 - 30)和104±11(BMI>30)(20 - 25与>30相比,p<0.01)。消融前平均ECI值与BMI呈正相关(r = 0.50)。本研究记录了PV隔离期间ECI值的预期范围。观察到的ECI值与患者BMI相关。讨论了当前一代Contact系统的潜在局限性及未来临床应用前景。