Nellcor™ PM1000N与Masimo Radical-7用于检测志愿者呼吸暂停的比较。

Comparison of Nellcor™ PM1000N and Masimo Radical-7 for detecting apnea in volunteers.

作者信息

Kitsiripant Chanatthee, Fukada Tomoko, Iwakiri Hiroko, Tsuchiya Yuri, Ozaki Makoto, Nomura Minoru

机构信息

Department of Anesthesiology, School of Medicine, Tokyo Women's Medical University, 8-1 Kawada-cho Shinjuku-ku, Tokyo, 1628666, Japan.

Department of Anesthesiology, Faculty of Medicine, Prince of Songkla University, Songkhla, Thailand.

出版信息

J Anesth. 2017 Oct;31(5):709-713. doi: 10.1007/s00540-017-2385-4. Epub 2017 Jul 20.

Abstract

PURPOSE

Although capnography is considered the gold standard for monitoring of ventilation, it may not work accurately in some situations. We compared the performance of two non-invasive continuous respiratory rate (RR) monitors that are alternatives for the detection of respiratory depression.

METHODS

Fifty healthy volunteers ≥20 years old were enrolled in this study. After monitoring of the volunteers was started by the Masimo Radical-7 and Nellcor™ PM1000N, they breathed at the rate of 12 breaths/min for 3 min and then stopped breathing. As soon as the apnea alarm of either monitor went off, breathing was resumed at the same rate. This entire procedure was repeated three times. The data collected every 30 s included the RR, pulse rate (PR) and oxygen saturation (SpO). The times of alarm on, alarm off and reappearance of RR on the monitor screens were also recorded.

RESULTS

The biases of the RR, PR and SpO measurements from the two monitors were 0.5, 0.2 and -0.4, respectively. Of 143 procedures that stopped breathing for more than 30 s, 114 and 15 alarms were shown by the Masimo Radical-7 and Nellcor™ PM-1000N monitors, respectively. Most alarms of the Nellcor™ PM1000N followed from SpO <90%. Conversely, most alarms of the Masimo Radical-7 were caused by RR <10 breaths/min. Times to alarm on, alarm off and display of RR measured by the Masimo Radical-7 monitor were significantly shorter than the Nellcor™ PM-1000N monitor.

CONCLUSIONS

The Masimo Radical-7 monitor provides better detection of apnea in volunteers than the Nellcor™ PM-1000N.

摘要

目的

虽然二氧化碳描记法被认为是监测通气的金标准,但在某些情况下可能无法准确工作。我们比较了两种用于检测呼吸抑制的无创连续呼吸频率(RR)监测仪的性能。

方法

招募了50名年龄≥20岁的健康志愿者参与本研究。在通过Masimo Radical-7和Nellcor™ PM1000N对志愿者进行监测后,他们以每分钟12次呼吸的频率呼吸3分钟,然后停止呼吸。一旦任何一台监测仪的呼吸暂停警报响起,就以相同频率恢复呼吸。整个过程重复三次。每30秒收集的数据包括RR、脉搏率(PR)和血氧饱和度(SpO)。还记录了监测仪屏幕上警报响起、警报解除以及RR再次出现的时间。

结果

两台监测仪测量的RR、PR和SpO的偏差分别为0.5、0.2和 -0.4。在143次呼吸停止超过30秒的过程中,Masimo Radical-7和Nellcor™ PM - 1000N监测仪分别显示了114次和15次警报。Nellcor™ PM1000N的大多数警报是由于SpO <90%。相反,Masimo Radical-7的大多数警报是由RR <10次/分钟引起的。Masimo Radical-7监测仪测量的警报响起、警报解除以及RR显示的时间明显短于Nellcor™ PM - 1000N监测仪。

结论

与Nellcor™ PM - 1000N相比,Masimo Radical-7监测仪在志愿者中对呼吸暂停的检测效果更好。