Orthopaedic Department, University Hospital Jena, Campus Eisenberg, Eisenberg, Germany.
Otto Schott Institute of Materials Research, Friedrich Schiller University Jena, Jena, Germany.
J Arthroplasty. 2017 Dec;32(12):3802-3805. doi: 10.1016/j.arth.2017.07.027. Epub 2017 Jul 25.
The most common reason for revision total hip arthroplasty remains polyethylene wear. Development dysplasia of the hip and revision situations requires a conscious compromise of implant position. The surgeon should know about the consequence on wear via a possible change in hip contact force. The objective of this study is to investigate whether annual wear is dependent on hip contact force.
Forty-five inserts (DuraLoc, DePuy) that were explanted in our department were included. Three-dimensional gravimetric determination of the wear was performed by fluid displacement. Then, the hip contact force was determined using radiographs according to the Blumentritt model.
No correlation was found between patient-specific factors and the annual wear. The hip contact force estimated by the Blumentritt model also showed no correlation between hip contact force and annual wear. Two single model parameters correlated significantly with wear: VRECAB as a ratio of the lever length of the spinocrural and the pelvitrochanteric muscles and the angle Alpha as a measure of the position of the center of rotation in relation to the greater trochanter. The greater the ratio spinocrural/pelvitrochanteric lever arm (R = 0.408, P = .005) and the greater the Alpha angle (more valgus the femoral neck) (R = 0.377, P = .011) were, the greater was the wear.
These results lead to the conclusion that neither patient-specific factors nor the estimated hip contact force have a major influence on annual wear in the case of DuraLoc cups. Only a coxa valga and a small femoral offset contribute in a limited amount to an increase in wear.
翻修全髋关节置换术最常见的原因仍然是聚乙烯磨损。髋关节发育不良和翻修情况需要有意识地妥协植入物的位置。外科医生应该了解通过髋关节接触力的可能变化对磨损的影响。本研究的目的是研究每年的磨损是否取决于髋关节接触力。
共纳入 45 个(DuraLoc,DePuy)植入物。采用流体置换法进行三维重力磨损测定。然后,根据 Blumentritt 模型通过 X 射线确定髋关节接触力。
患者特定因素与年磨损量之间未发现相关性。Blumentritt 模型估计的髋关节接触力与年磨损量之间也无相关性。两个单一模型参数与磨损显著相关:VRECAB 作为螺旋状和骨盆状肌肉杠杆长度的比值和 Alpha 角作为旋转中心相对于大转子位置的度量。螺旋状/骨盆状杠杆臂的比值越大(R = 0.408,P =.005)和 Alpha 角越大(股骨颈越外翻)(R = 0.377,P =.011),磨损越大。
这些结果表明,在 DuraLoc 杯的情况下,患者特定因素和估计的髋关节接触力对年磨损量没有重大影响。只有股骨颈外翻和小股骨偏移量在一定程度上增加了磨损量。