The Spectra Optia continuous mononuclear cell (CMNC apheresis) system has emerged as the preferred device in peripheral blood stem cell collections over the original two-step Spectra Optia mononuclear cell (MNC apheresis) system. Until now, no comparative data were available for non-stimulated MNC collections that are required for immunotherapy.
We compared collection parameters and product composition for Spectra Optia MNC- as well as CMNC-apheresis systems in non-stimulated MNC collections from 35 registry donors intended for donor lymphocyte infusions. In a subsequent analysis, different centrifugation forces (determined as packing factor or PF) were investigated regarding target cell yield and contamination in 61 collections using the CMNC device only.
Comparable collection efficiencies as well as target cell yields could be achieved with the Spectra Optia MNC- versus CMNC program. Similar numbers of MNC, T, B and NK cells could be collected with both devices. This led to a more than twofold lymphocyte recruitment from lymphatic tissue into the blood during apheresis. However, significantly more blood had to be processed with longer procedure time using the MNC program resulting in larger product volumes compared to the CMNC setting. Red blood cell and platelet (PLT) contamination were similar. Lowering the centrifugation force from PF4·5 to PF4·0 significantly reduced PLT contamination without affecting target cell yield in the product.
The Spectra Optia CMNC device using lower centrifugal force (PF4·0) showed similar target cell yield and composition as well as collection efficiencies with superior performance parameters and lower PLT contamination compared to the MNC setting.
相较于最初的两步式Spectra Optia单核细胞采集系统,Spectra Optia连续单核细胞采集系统(CMNC单采)已成为外周血干细胞采集的首选设备。到目前为止,尚无关于免疫治疗所需非刺激单核细胞采集的比较数据。
我们比较了Spectra Optia单核细胞采集系统以及CMNC采集系统在35名登记捐献者非刺激单核细胞采集中的采集参数和产品组成,这些捐献者用于供体淋巴细胞输注。在后续分析中,仅使用CMNC设备对61次采集中不同的离心力(以压实因子或PF确定)进行了目标细胞产量和污染情况的研究。
Spectra Optia单核细胞采集程序与CMNC采集程序的采集效率和目标细胞产量相当。两种设备采集到的单核细胞、T细胞、B细胞和NK细胞数量相似。这导致在单采过程中从淋巴组织募集到血液中的淋巴细胞数量增加了两倍多。然而,与CMNC设置相比,使用单核细胞采集程序需要处理更多的血液,且程序时间更长,从而导致产品体积更大。红细胞和血小板污染情况相似。将离心力从PF4·5降至PF4·0可显著降低血小板污染,且不影响产品中的目标细胞产量。
与单核细胞采集设置相比,使用较低离心力(PF4·0)的Spectra Optia CMNC设备显示出相似的目标细胞产量、组成和采集效率,且性能参数更优,血小板污染更低。